04.01.2015 Views

CR1000 Manual - Campbell Scientific

CR1000 Manual - Campbell Scientific

CR1000 Manual - Campbell Scientific

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Section 7. Installation<br />

7.8.1.6.1 Quarter-Bridge Shunt (Option 13)<br />

With CRBasic example FieldCalStrain() Calibration Demo (p. 164) sent to the<br />

<strong>CR1000</strong>, and the strain gage stable, use the external keyboard / display or<br />

software numeric monitor to change the value in variable KnownRes to the<br />

nominal resistance of the gage, 1000 Ω, as shown in figure Strain-Gage Shunt<br />

Calibration Started (p. 165). Set Shunt_Mode to 1 to start the two-point shunt<br />

calibration. When Shunt_Mode increments to 3, the first step is complete.<br />

To complete the calibration, shunt R1 with the 249-kΩ resistor. Set variable<br />

KnownRes to 249000. As shown in figure Strain-Gage Shunt Calibration<br />

Finished (p. 165), set Shunt_Mode to 4. When Shunt_Mode = 6, shunt calibration<br />

is complete.<br />

Figure 50: Strain-gage shunt calibration started<br />

Figure 51: Strain-gage shunt calibration finished<br />

7.8.1.6.2 Quarter-Bridge Zero (Option 10)<br />

Continuing from Quarter-Bridge Shunt (Option 13) (p. 165), keep the 249-kΩ<br />

resistor in place to simulate a strain. Using the external keyboard / display or<br />

software numeric monitor, change the value in variable Zero_Mode to 1 to start<br />

the zero calibration as shown in figure Starting Zero Procedure (p. 166). When<br />

Zero_Mode increments to 6, zero calibration is complete as shown in figure Zero<br />

Procedure Finished (p. 166).<br />

165

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!