04.01.2015 Views

CR1000 Manual - Campbell Scientific

CR1000 Manual - Campbell Scientific

CR1000 Manual - Campbell Scientific

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Section 7. Installation<br />

"offset" = "y‐ intercept" = "zero"<br />

"multiplier" = "slope" = "gain"<br />

7.8.1.5.1 Zero or Tare (Option 0)<br />

Zero option simply adjusts a sensor's output to zero. It does not affect the<br />

multiplier.<br />

Case: A sensor measures the relative humidity (RH) of air. Multiplier is known to<br />

be stable, but sensor offset drifts and requires regular zeroing in a desiccated<br />

chamber. The following procedure zeros the RH sensor to obtain the calibration<br />

report shown. Use the external keyboard / display or software numeric monitor to<br />

change variable values as directed.<br />

Table 22. Calibration Report for Air RH Sensor<br />

Parameter<br />

Argument at Deployment<br />

mV output 1000 mV 1050 mV<br />

KnownRH (desiccated<br />

chamber)<br />

0 % 0 %<br />

Multiplier 0.05 % / mV 0.05 % / mV<br />

Offset -50 % -52.5 %<br />

RH reading 0 % 0 %<br />

Argument at 30-Day<br />

Service<br />

1. Send CRBasic example FieldCal Zeroing Demonstration Program (p. 155) to the<br />

<strong>CR1000</strong>. An excitation channel has been programmed to simulate a sensor<br />

output.<br />

2. To place the simulated RH sensor in a simulated-calibration condition (in the<br />

field it would be placed in a desiccated chamber), place a jumper wire between<br />

channels VX1/EX1 and SE6 (3L). Set variable mV to 1000. Set variable<br />

KnownRH to 0.0.<br />

3. To simulate a calibration, change the value in variable CalMode to 1 to start<br />

calibration. When CalMode increments to 6, zero calibration is complete.<br />

Calibrated Offset will equal -50% at this stage of this example.<br />

Figure 48: Zero (Option 0)<br />

4. To continue this example and simulate a zero-drift condition, change variable<br />

mV to 1050.<br />

154

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!