04.01.2015 Views

CR1000 Manual - Campbell Scientific

CR1000 Manual - Campbell Scientific

CR1000 Manual - Campbell Scientific

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Section 7. Installation<br />

Mode Variable<br />

Interpretation<br />

> 0 and ≠ 6 calibration in progress<br />

< 0 calibration encountered an error<br />

2 calibration in process<br />

6 calibration complete.<br />

7.8.1.4.2 Two-point Calibrations (multiplier / gain)<br />

Use this two-point calibration procedure to adjust multipliers (slopes) and offsets<br />

(y-intercepts). See Two Point Slope and Offset (Option 2) (p. 159) and Two Point<br />

Slope Only (Option 3) (p. 161) for demonstration programs:<br />

1. Use a separate FieldCal() instruction and separate, related variables for each<br />

sensor to be calibrated.<br />

2. Ensure mode variable = 0 or 6 before starting.<br />

a. If Mode > 0 and ≠ 6, calibration is in progress.<br />

b. If Mode < 0, calibration encountered an error.<br />

3. Place sensor into first known point condition.<br />

4. Set KnownVar variable to first known point.<br />

5. Set Mode variable = 1 to start first part of calibration.<br />

a. Mode = 2 (automatic) during the first point calibration.<br />

b. Mode = 3 (automatic) when the first point is completed.<br />

6. Place sensor into second known point condition.<br />

7. Set KnownVar variable to second known point.<br />

8. Set Mode = 4 to start second part of calibration.<br />

a. Mode = 5 (automatic) during second point calibration.<br />

b. Mode = 6 (automatic) when calibration is complete.<br />

7.8.1.5 FieldCal() Demonstration Programs<br />

FieldCal() has the following calibration options:<br />

• Zero<br />

• Offset<br />

• Two-point slope and offset<br />

• Two-point slope only<br />

• Zero basis (multi-put zero)<br />

These demonstration programs are provided as an aid in becoming familiar with<br />

the FieldCal() features at a test bench without actual sensors. For the purpose of<br />

the demonstration, sensor signals are simulated by the <strong>CR1000</strong> excitation channel.<br />

To reset tests, use the support software File Control (p. 454) menu commands to<br />

delete .cal files, and then send the demonstration program again to the <strong>CR1000</strong>.<br />

Term equivalents are as follows:<br />

153

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!