4.2.4 Migration time/mobility reproducibility<br />

Run-to-run migration time and mobility reproducibility in CE<br />

can often be better than 0.5 % RSD. This value is dependent<br />

on the condition of <strong>capillary</strong> wall, the composition, pH, and<br />

viscosity of the buffer, and the nature of the sample, and<br />

quality of instrumentation. Additional detail is given in table<br />

17.<br />

Factor Causeleffect Solution<br />

Temperature • Changes viscosity • Thermostat <strong>capillary</strong><br />

change<br />

and EOF<br />

Adsorption to • Changes EOF • Condition <strong>capillary</strong><br />

<strong>capillary</strong> walls • Caused by buffer, additive, and allow sufficient<br />

or sample adsorption equilibration time<br />

Hysteresis of • Caused by conditioning • Avoid pH differences<br />

wall charge<br />

<strong>capillary</strong> at high (or low)<br />

pH and employing a low • Allow sufficient<br />

(or high) pH running buffer equilibration time<br />

Changes in buffe • pH changes due to • Replenish buffer<br />

composition<br />

electrolysis<br />

• Buffer evaporation<br />

• Cap buffer vials and<br />

cool carousel<br />

• Conditioning waste<br />

• Use separate reservoir<br />

flushed into outlet<br />

to collect wash solutions<br />

reservoir<br />

• Carrying sodium hydroxide • First dip <strong>capillary</strong> in<br />

from conditioning vial into separate buffer or<br />

buffer vial, for example water vial<br />

Buffer reservoirs • Non-reproducible • Level liquid in reservoirs<br />

not level laminar flow • If not replenishing, do<br />

not use inlet vial for<br />

washing <strong>capillary</strong><br />

Different silanol • Different wall charge • Measure EOF and<br />

content of silica and variations in EOF normalize if necessary<br />

batches<br />

Instrumentation/Operation<br />

Table 17<br />

Factors affecting migration time<br />

reproducibility<br />

Variations in • Proportional changes • Not user accessible<br />

applied voltage in migration time<br />

95