High performance capillary electrophoresis - T.E.A.M.
High performance capillary electrophoresis - T.E.A.M.
High performance capillary electrophoresis - T.E.A.M.
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Modes<br />
EOF [mm/min]<br />
50<br />
40<br />
30<br />
20<br />
10<br />
These covalent modifications are intended to be permanent<br />
and to require little or no maintenance. Since the capillaries<br />
are usually washed after use (adsorption may occur even<br />
with the coating), they must be stable to washing solutions<br />
and to hydrodynamic flow. Unfortunately, the stability of<br />
most coatings is limited. It is anticipated that numerous<br />
types of stable coatings will soon be purchasable, similar to<br />
LC and GC columns.<br />
0<br />
-10<br />
-20<br />
0 2 4 6 8 10<br />
pH<br />
Figure 26<br />
Reversible electroosmotic flow in a<br />
a-lactalbumin-coated <strong>capillary</strong> 12<br />
3.1.2.2 Dynamic deactivation<br />
Addition of modifiers to the running buffer is an alternative<br />
to the bonded or adhered phases. An advantage of dynamic<br />
coatings is stability. Since the modifier is in the buffer, the<br />
coating is continuously regenerated and permanent stability<br />
is not required.<br />
As with covalent coatings, additives can interact with the<br />
wall and alter charge and hydrophobicity. These modifiers<br />
are both easier to implement and optimize since they are<br />
prepared by simple dissolution of the modifier in the buffer.<br />
Several dynamic deactivation methods are listed in table 13<br />
and an illustration of the use of cationic surfactants to<br />
reverse the EOF was shown in figure 22.<br />
A potential disadvantage of the dynamic modification<br />
approach is that solutes as well as the <strong>capillary</strong> surface are<br />
affected. Biological-type conditions will be sacrificed by the<br />
use of pH extremes and addition of surfactants. Another<br />
limitation can be the equilibration time needed to obtain a<br />
reproducible surface and constant EOF. Furthermore, postcolumn<br />
analyses such as mass spectrometry and enzymatic<br />
assays are sensitive to additives, especially those in high<br />
concentrations.<br />
57