High performance capillary electrophoresis - T.E.A.M.

High performance capillary electrophoresis - T.E.A.M. High performance capillary electrophoresis - T.E.A.M.

03.01.2015 Views

Instrumentation/Operation mAU 140 120 100 80 60 40 20 140 120 100 80 60 40 20 0 Figure 67 Improved migration time reproducibility using buffer replenishment Overlay of runs 5, 10, and 15 a) replenish after every 5 runs b) no replenishment a) 3.0 3.5 4.0 4.5 5.0 5.5 b) 3.0 3.5 4.0 4.5 5.0 5.5 Time [min] system, the volume of the reservoirs, and whether the capillary conditioning washes are flushed into the exit reservoir (not recommended). For these reasons, frequent replacement of buffer is recommended. An example of the benefit of frequent buffer replenishment is shown in figure 67. The replenishment system operates by emptying the contents of the reservoirs into a waste bottle and then refilling them with fresh buffer. With a large buffer replenishment system minimal vial locations in the autosampler are needed for buffer, leaving more locations for samples. In addition, a large reservoir affords the capability to run automated analysis for a long period of time (over the weekend, for example). 4.4.4 Buffer leveling As described in chapter 2, maintaining equal liquid levels is important to both efficiency and migration time reproducibility. Siphoning from one reservoir to the other will superimpose laminar flow into the system. This is not only detrimental to reproducibility but also to the efficiency. The influence of non-level reservoirs is dependent on the diameter and length of the capillary and the viscosity of the buffer. Clearly, wide-bore, short capillaries run at elevated temperatures would exhibit the most deleterious effects. It has been shown that a height difference of 2 mm in a 50-µm id capillary can change migration time 2 to 3 %. This increases to about 10 % in a 100-µm id capillary. An automated buffer leveling system would accurately deliver liquid to a user-specified level during replenishment. It can also be used to re-level liquids in the running reservoirs caused by EOF, without emptying them first. 112

Chapter 5 List of abbreviations 113

Instrumentation/Operation<br />

mAU<br />

140<br />

120<br />

100<br />

80<br />

60<br />

40<br />

20<br />

140<br />

120<br />

100<br />

80<br />

60<br />

40<br />

20<br />

0<br />

Figure 67<br />

Improved migration time reproducibility<br />

using buffer replenishment<br />

Overlay of runs 5, 10, and 15<br />

a) replenish after every 5 runs<br />

b) no replenishment<br />

a)<br />

3.0 3.5 4.0 4.5 5.0 5.5<br />

b)<br />

3.0 3.5 4.0 4.5 5.0 5.5<br />

Time [min]<br />

system, the volume of the reservoirs, and whether the<br />

<strong>capillary</strong> conditioning washes are flushed into the exit<br />

reservoir (not recommended). For these reasons, frequent<br />

replacement of buffer is recommended. An example of the<br />

benefit of frequent buffer replenishment is shown in figure<br />

67.<br />

The replenishment system operates by emptying the contents<br />

of the reservoirs into a waste bottle and then refilling<br />

them with fresh buffer. With a large buffer replenishment<br />

system minimal vial locations in the autosampler are needed<br />

for buffer, leaving more locations for samples. In addition,<br />

a large reservoir affords the capability to run automated<br />

analysis for a long period of time (over the weekend, for<br />

example).<br />

4.4.4 Buffer leveling<br />

As described in chapter 2, maintaining equal liquid levels is<br />

important to both efficiency and migration time reproducibility.<br />

Siphoning from one reservoir to the other will superimpose<br />

laminar flow into the system. This is not only detrimental<br />

to reproducibility but also to the efficiency. The<br />

influence of non-level reservoirs is dependent on the diameter<br />

and length of the <strong>capillary</strong> and the viscosity of the<br />

buffer. Clearly, wide-bore, short capillaries run at elevated<br />

temperatures would exhibit the most deleterious effects.<br />

It has been shown that a height difference of 2 mm in a<br />

50-µm id <strong>capillary</strong> can change migration time 2 to 3 %. This<br />

increases to about 10 % in a 100-µm id <strong>capillary</strong>.<br />

An automated buffer leveling system would accurately<br />

deliver liquid to a user-specified level during replenishment.<br />

It can also be used to re-level liquids in the running reservoirs<br />

caused by EOF, without emptying them first.<br />

112

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!