High performance capillary electrophoresis - T.E.A.M.
High performance capillary electrophoresis - T.E.A.M. High performance capillary electrophoresis - T.E.A.M.
Instrumentation/Operation mAU 140 120 100 80 60 40 20 140 120 100 80 60 40 20 0 Figure 67 Improved migration time reproducibility using buffer replenishment Overlay of runs 5, 10, and 15 a) replenish after every 5 runs b) no replenishment a) 3.0 3.5 4.0 4.5 5.0 5.5 b) 3.0 3.5 4.0 4.5 5.0 5.5 Time [min] system, the volume of the reservoirs, and whether the capillary conditioning washes are flushed into the exit reservoir (not recommended). For these reasons, frequent replacement of buffer is recommended. An example of the benefit of frequent buffer replenishment is shown in figure 67. The replenishment system operates by emptying the contents of the reservoirs into a waste bottle and then refilling them with fresh buffer. With a large buffer replenishment system minimal vial locations in the autosampler are needed for buffer, leaving more locations for samples. In addition, a large reservoir affords the capability to run automated analysis for a long period of time (over the weekend, for example). 4.4.4 Buffer leveling As described in chapter 2, maintaining equal liquid levels is important to both efficiency and migration time reproducibility. Siphoning from one reservoir to the other will superimpose laminar flow into the system. This is not only detrimental to reproducibility but also to the efficiency. The influence of non-level reservoirs is dependent on the diameter and length of the capillary and the viscosity of the buffer. Clearly, wide-bore, short capillaries run at elevated temperatures would exhibit the most deleterious effects. It has been shown that a height difference of 2 mm in a 50-µm id capillary can change migration time 2 to 3 %. This increases to about 10 % in a 100-µm id capillary. An automated buffer leveling system would accurately deliver liquid to a user-specified level during replenishment. It can also be used to re-level liquids in the running reservoirs caused by EOF, without emptying them first. 112
Chapter 5 List of abbreviations 113
- Page 62 and 63: Modes Figure 33 Ion analysis of fer
- Page 64 and 65: Modes The separation mechanism of n
- Page 66 and 67: Modes the stationary phase in LC. S
- Page 68 and 69: Modes Amplitude 2 a) with a migrati
- Page 70 and 71: Modes CGE t = 0 t > 0 Polymer matri
- Page 72 and 73: Modes Crosslinked polyacrylamide, a
- Page 74 and 75: Modes a) ds 500 base pairs This sam
- Page 76 and 77: Modes and resolution with respect t
- Page 78 and 79: Modes 3.5 Capillary isotachophoresi
- Page 80 and 81: Modes 80
- Page 82 and 83: Instrumentation/Operation Diode-arr
- Page 84 and 85: Instrumentation/Operation Pressure
- Page 86 and 87: Instrumentation/Operation If sensit
- Page 88 and 89: Instrumentation/Operation Despite q
- Page 90 and 91: Instrumentation/Operation T, ˚C 10
- Page 92 and 93: Instrumentation/Operation 4.2.1.1 C
- Page 94 and 95: Instrumentation/Operation However,
- Page 96 and 97: Instrumentation/Operation Calculati
- Page 98 and 99: Instrumentation/Operation Method Ma
- Page 100 and 101: Instrumentation/Operation 4.3.3 Lin
- Page 102 and 103: Instrumentation/Operation Area (arb
- Page 104 and 105: Instrumentation/Operation 4.3.6 Ext
- Page 106 and 107: Instrumentation/Operation light int
- Page 108 and 109: Instrumentation/Operation 4.3.7.2 Q
- Page 110 and 111: Instrumentation/Operation the capab
- Page 114 and 115: Abbreviations Abbreviation Full Nam
- Page 116 and 117: Abbreviations Abbreviation Full Nam
- Page 118 and 119: Abbreviations Abbreviation Full Nam
- Page 120 and 121: References/bibliography References
- Page 122 and 123: References/bibliography 15 W.C. Bru
- Page 124 and 125: References/bibliography 29 R.L. Chi
- Page 126 and 127: References/bibliography 126
- Page 128 and 129: Index A Absorption. See Detection,
- Page 130 and 131: Index D DAD. See Detection, diode-a
- Page 132 and 133: Index I Ionic strength , 22, 25, 26
- Page 134 and 135: Index selectivity, 47 Response time
- Page 136: www.agilent.com/chem Copyright © 2
Instrumentation/Operation<br />
mAU<br />
140<br />
120<br />
100<br />
80<br />
60<br />
40<br />
20<br />
140<br />
120<br />
100<br />
80<br />
60<br />
40<br />
20<br />
0<br />
Figure 67<br />
Improved migration time reproducibility<br />
using buffer replenishment<br />
Overlay of runs 5, 10, and 15<br />
a) replenish after every 5 runs<br />
b) no replenishment<br />
a)<br />
3.0 3.5 4.0 4.5 5.0 5.5<br />
b)<br />
3.0 3.5 4.0 4.5 5.0 5.5<br />
Time [min]<br />
system, the volume of the reservoirs, and whether the<br />
<strong>capillary</strong> conditioning washes are flushed into the exit<br />
reservoir (not recommended). For these reasons, frequent<br />
replacement of buffer is recommended. An example of the<br />
benefit of frequent buffer replenishment is shown in figure<br />
67.<br />
The replenishment system operates by emptying the contents<br />
of the reservoirs into a waste bottle and then refilling<br />
them with fresh buffer. With a large buffer replenishment<br />
system minimal vial locations in the autosampler are needed<br />
for buffer, leaving more locations for samples. In addition,<br />
a large reservoir affords the capability to run automated<br />
analysis for a long period of time (over the weekend, for<br />
example).<br />
4.4.4 Buffer leveling<br />
As described in chapter 2, maintaining equal liquid levels is<br />
important to both efficiency and migration time reproducibility.<br />
Siphoning from one reservoir to the other will superimpose<br />
laminar flow into the system. This is not only detrimental<br />
to reproducibility but also to the efficiency. The<br />
influence of non-level reservoirs is dependent on the diameter<br />
and length of the <strong>capillary</strong> and the viscosity of the<br />
buffer. Clearly, wide-bore, short capillaries run at elevated<br />
temperatures would exhibit the most deleterious effects.<br />
It has been shown that a height difference of 2 mm in a<br />
50-µm id <strong>capillary</strong> can change migration time 2 to 3 %. This<br />
increases to about 10 % in a 100-µm id <strong>capillary</strong>.<br />
An automated buffer leveling system would accurately<br />
deliver liquid to a user-specified level during replenishment.<br />
It can also be used to re-level liquids in the running reservoirs<br />
caused by EOF, without emptying them first.<br />
112