Section 4 - Incoming Acceptance<br />
Set the oscillator to 100kHz and the meter should read 220mV or less (–50dB).<br />
Set rear panel differential switch to `+’ and turn [ALL CH] mode on.<br />
4.8 DISTORTION AND MAXIMUM SIGNAL CHECKS<br />
Set the filter to a cutoff frequency of 5kHz in the low-pass mode with 0dB of Input and Output gain.<br />
Connect a low distortion oscillator to the INPUT and apply a 1Vrms signal at 500Hz.<br />
CAUTION<br />
If the distortion is excessive, verify that the distortion<br />
of the oscillator being used is