Shift Test<br />
This test involves moving a weight off the center point of the Top Plate to check for<br />
continued accuracy.<br />
1. Place an appropriate Test Load weight for each weight feature on each of the<br />
four circles individually. The display must show the same weight between any of<br />
the 4 readings.<br />
2. Remove all weights. The display must read 0.000 ± 0.005 kg (0.00 ± 0.00l0b)<br />
Test max thirty pounds (30.00 lb) in center only.<br />
3. Press and hold the <strong>Scale</strong> Zero button. Record the Cal and Par values shown on<br />
the display.