Shift Test<br />

This test involves moving a weight off the center point of the Top Plate to check for<br />

continued accuracy.<br />

1. Place an appropriate Test Load weight for each weight feature on each of the<br />

four circles individually. The display must show the same weight between any of<br />

the 4 readings.<br />

2. Remove all weights. The display must read 0.000 ± 0.005 kg (0.00 ± 0.00l0b)<br />

Test max thirty pounds (30.00 lb) in center only.<br />

3. Press and hold the <strong>Scale</strong> Zero button. Record the Cal and Par values shown on<br />

the display.