NCR 7878 Scale Calibration - ScanSource Technical Services
NCR 7878 Scale Calibration - ScanSource Technical Services
NCR 7878 Scale Calibration - ScanSource Technical Services
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
After calibrating the scale, the accuracy must be verified to assure it meets the<br />
government requirements. This consists of a series of four tests that must be run in the<br />
continuous sequence given.<br />
1. Increasing Load Test<br />
2. Over-Capacity Test<br />
3. Decreasing Load Test<br />
4. Shift Test<br />
Increasing Load Test<br />
This test checks the scale’s accuracy when incrementally adding weight to the center of<br />
the Top Plate. Use weights that correspond to the RealScan 78 weight features.<br />
Note: Do NOT remove any weight from the Top Plate between test steps unless<br />
directed to do so.