After calibrating the scale, the accuracy must be verified to assure it meets the<br />

government requirements. This consists of a series of four tests that must be run in the<br />

continuous sequence given.<br />

1. Increasing Load Test<br />

2. Over-Capacity Test<br />

3. Decreasing Load Test<br />

4. Shift Test<br />

Increasing Load Test<br />

This test checks the scale’s accuracy when incrementally adding weight to the center of<br />

the Top Plate. Use weights that correspond to the RealScan 78 weight features.<br />

Note: Do NOT remove any weight from the Top Plate between test steps unless<br />

directed to do so.