09.11.2012 Views

Fluid Jetting for Next Generation Packages - Nordson ASYMTEK 首页

Fluid Jetting for Next Generation Packages - Nordson ASYMTEK 首页

Fluid Jetting for Next Generation Packages - Nordson ASYMTEK 首页

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

the DJ-2000 the best dots of Loctite SMA-<br />

3621 and of the Silver Paste occurred at 0.12<br />

mm stroke length and at 0.15 mm stroke<br />

length, respectively. These were both the<br />

shortest stroke lengths that could be achieved<br />

(at lower stroke length dots would no longer<br />

be dispensed). At lower stroke lengths the<br />

spring is compressed less, so there is less<br />

<strong>for</strong>ce slamming the needle back to its seat.<br />

Because of the consistency of the underfill (it<br />

is similar to syrup or honey), perhaps greater<br />

<strong>for</strong>ce is needed to separate a drop from the<br />

rest of the fluid. Even with 4 turns of preload<br />

the stroke length may need to be longer to<br />

get the required compression of the spring in<br />

order to jet a dot. The final parameter<br />

affecting the size of the dots was the<br />

temperature the fluid was heated to. It was<br />

found that by adjusting the temperature and<br />

leaving all other setting constant the dot size<br />

was not changed. However, when the fluid<br />

was heated to higher temperatures, the stroke<br />

length could be set lower be<strong>for</strong>e it began to<br />

clog. There<strong>for</strong>e, with more heat, the stroke<br />

length could be set lower, giving smaller<br />

dots, so temperature indirectly affected dot<br />

size. The other adjustable parameters did not<br />

seem to influence dot size quite as much. At<br />

higher preloads there is more <strong>for</strong>ce slamming<br />

the needle into the seat. This causes dots to<br />

be jetted at a higher velocity. With a higher<br />

velocity, the drops have greater momentum<br />

when they impact the surface and tend to<br />

spread out and flatten more. The amount of<br />

preload had to be balanced, so that there was<br />

enough <strong>for</strong>ce to eject a drop but not so much<br />

to flatten out the dot. However, the effect of<br />

this on the dot size was minimal (even at<br />

lower preloads the dots were still being<br />

ejected at a high velocity). Pressure was<br />

another parameter that had to be balanced.<br />

Pressure needed to be high enough to <strong>for</strong>ce<br />

fluid below and around the needle when it<br />

was raised, but also not so high that the<br />

pressure to back pressure difference was too<br />

large. With a pressure much greater than the<br />

backpressure, a greater amount of fluid was<br />

jetted with each drop. However, even when<br />

Pac Tech, Berlin, April 2002<br />

varying pressure between 10 and 50 psi, the<br />

influence of this effect on dot size was small<br />

compared to the influence of other<br />

parameters. The final adjustment made was<br />

the switch between the 4 and 5 mil nozzles.<br />

It was observed that the dots produced by<br />

either nozzle were comparable in size, and<br />

the 4-mil nozzle clogged more often than the<br />

5-mil nozzle. Because the nozzle size did not<br />

appear to affect dot size and have the<br />

clogging problems with the 4-mil nozzle, the<br />

5-mil nozzle was used <strong>for</strong> the rest of the<br />

testing. A final observation regarding the<br />

variability of dot size based on parameter<br />

settings is the potential benefit of this. The<br />

ability to adjust a few parameters and get<br />

dots which range in size from 12 mils<br />

diameter to greater than 30 mil in diameter,<br />

along with the ability to fire multiple dots in<br />

the same place (to get bigger dots) makes the<br />

DJ-2000 an extremely versatile piece of<br />

equipment.<br />

Recommendations:<br />

During the course of the testing several<br />

observations were made that could lead to<br />

improvements in the current DJ-2000 design.<br />

Currently the needle is stopped in its<br />

downward descent by slamming into the<br />

seat, which absorbs all of the <strong>for</strong>ce, and<br />

there<strong>for</strong>e must be constructed of carbide to<br />

withstand this wear. This extra cost (<strong>for</strong> the<br />

carbide parts) could be avoided as follows.<br />

By integrating an adjustable ring inside of<br />

the tube, and by adding a ridge on the shaft<br />

of the needle, the <strong>for</strong>ce could be transferred<br />

from the seat. The ring could be set so that<br />

the ridge on the needle comes to rest on the<br />

top of the ring. This would provide the<br />

stopping surface <strong>for</strong> the needle, rather than<br />

the seat. Because the contact area of this ring<br />

would be much greater than that of the seat,<br />

the <strong>for</strong>ce would be spread out more, and the<br />

component would not need to be made of<br />

carbide. The second proposed modification is<br />

to adjust the seat design to increase the<br />

backpressure on the fluid (the pressure<br />

pushing it up into the syringe). Having

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!