Maintenance Instructions<br />
UL CIRUIT CALIBRATION (continued)<br />
NOTE: Certified weights can be added to check positive response and then<br />
removed to check negative response.<br />
7. Reassemble oil bottle and shoud back on the unit.<br />
8. While the circuit calibration is now finished, the scale must still be calibrated<br />
to complete the calibrating procedure.<br />
* The P-1 POT (Potentiometer) is located below the keypad at the front of the<br />
circuit board.<br />
NOTE: This entire procedure must be completed to update the scale calibration<br />
memory of the unit. turning off the power at any time dureing the procedure returns<br />
the unit to the previous calibration values and the procedure must be started again<br />
from the beginning.<br />