Technical Data<br />

UL (Underwr<strong>it</strong>ers Laboratories Inc.)<br />

CSA (Canadian Standards Association)<br />

P3S<br />

Only as to Intrinsic Safety for use in Hazardous Locations<br />

Class I, Div. 1, Groups A, B, C, D<br />

Class II, Div. 1, Groups E, F, G<br />

Use in accordance w<strong>it</strong>h <strong>Dräger</strong> Control Drawing SE20105.<br />

T4: –40 ≤ Ta ≤ +65 o C, T6: –40 ≤ Ta ≤ +40 o C.<br />

Not tested in oxygen enriched atmospheres (>21 % O 2 ).<br />

Power Supply: V max = 30 V, I max = 0.3 A, P i = 700 mW,<br />

C i = 0 nF, L i = 50 μH<br />

Meter circu<strong>it</strong>: V oc = 7.6 V, I sc = 1 mA, V max = 10.4 V, C a = 2.5 μF, L a = 10 mH,<br />

C i = 0, L i = 0<br />

P3S<br />

Intrinsic safe<br />

®<br />

Class I, Div. 1, Groups A, B, C, D<br />

LR97594<br />

Class II, Div. 1, Groups E, F, G<br />

Ex ia T4 (–40 o C Ta +65 o NRTL/C<br />

C),<br />

Ex ia T6 (–40 o C Ta +40 o C)<br />

Signal transmission to central un<strong>it</strong><br />

Analogue<br />

— Measured-value signal 4 mA to 20 mA<br />

— Drift below zero point 3.8 mA to 4 mA<br />

— Full-scale value exceeded 20 mA to 20.5 mA<br />

— Un<strong>it</strong> fault