MS<strong>16A</strong> Syringe Driver<br />
Smiths Medical<br />
11. Sounder check<br />
Check that whenever power is restored (e.g. battery removed and<br />
replaced), an undistorted sound, clearly audible at arms length and<br />
lasting 5 - 20 seconds, is produced.<br />
12. General<br />
a. Check that both case halves are in good condition and that all<br />
interior screws are present and tightened.<br />
b. Remove the dummy case and fit the rear case (see Chapter 4,<br />
Case assembly).<br />
c. Affix new front label if required and seal unit.<br />
d. Clean the unit and fix the appropriate <strong>Service</strong> Label.<br />
e. Ensure that the History record sheet is completed.<br />
5<br />
5 — 10<br />
Issue 6 (February 2005)<br />
Technical <strong>Service</strong> <strong>Manual</strong>