Volts<br />

MS<strong>16A</strong> Syringe Driver<br />

Smiths Medical<br />

To measure the motor interval time, a conventional scope with a good<br />

trigger is required; a storage scope would be better. Using the<br />

variable power supply, place a 4R7 ohm resistor in the negative<br />

supply line, and with a x10 scope probe, measure the voltage across<br />

the resistor.<br />

Run the pump at maximum rate and compare each consecutive pulse<br />

over six pulses. Figure 5-3 illustrates the typical waveform expected.<br />

The motor run time is measured from the start of the plateau to the<br />

trailing edge of the waveform, where it is cut off by the cam’s action.<br />

0.16<br />

MS16 Motor Pulses<br />

0.14<br />

0.12<br />

0.10<br />

0.08<br />

0.06<br />

0.04<br />

0.02<br />

Cam Period<br />

5<br />

0<br />

0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 0.50 0.55 0.60 0.65<br />

GM0105_8044-GB-B<br />

Time (seconds)<br />

Figure 5-3 Typical waveform for one pulse<br />

The variation in duration of the six consecutive pulses in each<br />

revolution should not be more than 20%. If necessary, adjust the<br />

motor run-time ratio by a slight bending of the cam follower lever,<br />

using tool part number 0105-0079. Use the tool as shown in Figure 5-4<br />

to tweak the operating lever so as to mimimise the error.<br />

Note that the time interval between successive pulses may vary<br />

slightly. This is not a controlled parameter, and variation is<br />

permitted.<br />

5 — 6<br />

Issue 6 (February 2005)<br />

Technical <strong>Service</strong> <strong>Manual</strong>