MS<strong>16A</strong> Syringe Driver<br />

Smiths Medical<br />

Adjustment potentiometers<br />

The adjustment potentiometers VR5 and VR9 are behind the front<br />

panel label.<br />

Peel back the front panel label from the righthand side (nearest the<br />

LED) to expose the access holes to the resistors.<br />

Variable resistor VR9 (thrust) is located directly above the LED.<br />

VR5 is near the top.<br />

When adjustments have been completed, it may be necessary to<br />

replace the front panel label if it does not adhere when pressed back<br />

in position.<br />

When replacing the label, ensure that the small clear plastic window<br />

is in its recess over the rate dials.<br />

Notes: If any unit is found to be over or under infusing during<br />

testing, inform your local RAQA Department.<br />

Ensure that the button surround on the actuator body is blue<br />

(Figure 7-1, items 15 and 14)<br />