Smiths Medical<br />

MS<strong>16A</strong> Syringe Driver<br />

Removal and dismantling procedures<br />

1. Case separation<br />

a. Lay the driver face down on a clean flat surface.<br />

b. Slide off the battery cover (item 27) and if fitted, remove the<br />

battery.<br />

c. Remove the three screws and washer securing the halves of the<br />

case assembly. Lift off the rear half.<br />

d. If the case is sealed, gently pry the two halves apart. Remove any<br />

sealant from mating surfaces.<br />

2. Leadscrew and bearing assembly removal<br />

a. Separate the case halves as above.<br />

b. Loosen the motor clamp (item 13) so that the motor and gearbox<br />

assembly (item 14) can be tilted clear of the case.<br />

c. Gently pull the leadscrew and actuator assembly (items 15 and<br />

16) from the motor and gearbox assembly.<br />

3. Actuator and back bearing disassembly<br />

a. Separate the case halves and remove the leadscrew and bearing<br />

assembly as above.<br />

4<br />

4. PCB removal<br />

b. Remove the back bearing (item 20) from the end of the leadscrew.<br />

c. Depress the plunger (half nut - item 18) on the actuator assembly<br />

(item 19) and slide the actuator off the leadscrew at the back<br />

bearing end.<br />

When the half nut and spring are freed from the leadscrew,<br />

ensure that they are not mislaid. Ensure that the half nut is bluerimmed.<br />

a. Separate the case halves and remove the leadscrew and bearing<br />

assembly as above.<br />

b. Remove the motor clamp (item 13), screw and fibre washer.<br />

c. Remove the four remaining screws and fibre washers securing<br />

the PCB assembly (item 22) to the front case (item 2).<br />

If the wires for the piezo-transducer assembly (item 25) are<br />

secured with an adhesive pad to the motor clamp (item 13),<br />

carefully lift the leads from the pad.<br />

d. Pull the battery connector springs from their locating slots and<br />

lift out the PCB assembly and motor and gearbox assembly.<br />

Technical <strong>Service</strong> <strong>Manual</strong> Issue 6 (February 2005) 4 — 5