Smiths Medical<br />

MS<strong>16A</strong> Syringe Driver<br />

Chapter 4 - Disassembly and Assembly<br />

General<br />

Warnings:<br />

Refer all service, repair, testing and calibrations only to suitably<br />

qualified technical personnel. Unauthorised modifications to the<br />

driver must not be carried out.<br />

The use of spare parts other than those recommended is not advised,<br />

and may affect the correct operation of the driver, resulting in injury or<br />

death.<br />

Cautions:<br />

The moving parts of the driver do not require any lubrication during<br />

their lifetime. Any worn or stiff parts should be replaced.<br />

With the case dismantled take great care to avoid contamination of the<br />

exposed switches, e.g. by the entry of solder particles.<br />

Handling of printed circuit boards is required during disassembly/<br />

assembly. A static controlled work station including a conductive mat<br />

and grounded wrist strap should be used to provide protection<br />

against electrostatic discharge (ESD) or circuit board damage could<br />

result.<br />

4<br />

The following procedures describe the dismantling and assembly<br />

sequence when replacing parts.<br />

If the case halves are separated, the complete test routine (see<br />

Chapter 5, Testing) must be completed successfully.<br />

Unless otherwise stated, tighten screws to a torque of 30 - 35 cNm.<br />

A faulty PCB should be replaced with a new PCB.<br />

The item numbers given in the following descriptions refer to Figure<br />

4-1, MS<strong>16A</strong> exploded view.<br />

Tools, equipment and materials required<br />

Cam adjustment tool Part number 0105-0079<br />

Silicone grease Part number 6835-2006<br />

Silicone sealant, non-acetic Loctite 209161, or similar.<br />

For a list of additional tools and materials used when testing the<br />

MS<strong>16A</strong>, see Chapter 5, Testing, page 5-1.<br />

Technical <strong>Service</strong> <strong>Manual</strong> Issue 6 (February 2005) 4 — 3