Safety Devices to DIN12876<br />

- Low level switch<br />

- Adjustable over-temperature switch (also valid for units with heat<strong>in</strong>g)<br />

Classification of Laboratory Thermostats and Baths<br />

Classification Thermal Fluid Technical requirement Designation d<br />

I non-flammable a Over-temperature cut-off c NFL<br />

II Adjustable over-temperature cut-off<br />

flammable b FL<br />

III Adjustable over-temperature cut-off<br />

and extra low-level switch<br />

a Normally water; other fluids only when they are non-flammable <strong>in</strong> the event of a s<strong>in</strong>gle<br />

failure.<br />

b<br />

The thermal fluid must have a flame po<strong>in</strong>t ≥ 65 °C, this means that ethanol can only<br />

be used under constant supervision.<br />

c<br />

The over-temperature protection can for example be provided by a fluid sensor or<br />

a suitable over temperature switch.<br />

d Determ<strong>in</strong>ed by the manufacturer.<br />

Temperature control units with heat<strong>in</strong>g (marked by an "-H" <strong>in</strong> the name) are designated<br />

as class III FL.<br />

Temperature control units without heat<strong>in</strong>g are designated as class I NFL.<br />

Mechanical Over-temperature Switch<br />

This temperature control unit is equipped with a mechanical over-temperature switch.<br />

For sett<strong>in</strong>g the over-temperature protection refer to chapter sett<strong>in</strong>g the overtemperature<br />

switch.<br />

9