Display Error Messages<br />

Alarms and Warn<strong>in</strong>gs<br />

In case of malfunction alarms and warn<strong>in</strong>g messages are <strong>in</strong>dicated through the display.<br />

Display Cause Effect, measurement<br />

flash<strong>in</strong>g display Warn<strong>in</strong>g:<br />

Thermoregulation cont<strong>in</strong>ues.<br />

(temperature Over- undertemperature (limit<br />

value)<br />

value +/- 2K from set po<strong>in</strong>t).<br />

F1-<br />

malfunction sensor F1 Thermoregulation is <strong>in</strong>active<br />

flash<strong>in</strong>g<br />

broken or short circuit<br />

(Pump off, compressor off,<br />

heat<strong>in</strong>g off).<br />

Please check sensor.<br />

E1-<br />

flash<strong>in</strong>g<br />

E2-<br />

flash<strong>in</strong>g<br />

E3-<br />

flash<strong>in</strong>g<br />

EP-<br />

flash<strong>in</strong>g<br />

30<br />

Inlet E1 <strong>in</strong>dicates failure.<br />

No release-signal, level-alarm.<br />

Inlet E1 <strong>in</strong>dicates failure.<br />

Pump is runn<strong>in</strong>g and<br />

circulation is miss<strong>in</strong>g or Pump<br />

is runn<strong>in</strong>g and water pressure<br />

is miss<strong>in</strong>g.<br />

Even though thermoregulation<br />

is off, <strong>in</strong>let E1 is <strong>in</strong>dicat<strong>in</strong>g<br />

circulation.<br />

Loss of data <strong>in</strong> parameter<br />

memory.<br />

Thermoregulation is <strong>in</strong>active<br />

(Pump off, compressor off,<br />

heat<strong>in</strong>g off).<br />

Check level.<br />

Reset only possible if level ok.<br />

Thermoregulation is <strong>in</strong>active<br />

(Pump off, compressor off,<br />

heat<strong>in</strong>g off).<br />

Reset only possible if<br />

disconnected from the power<br />

supply.<br />

Thermoregulation is <strong>in</strong>active<br />

(Pump off, compressor off,<br />

heat<strong>in</strong>g off).<br />

Failure is be<strong>in</strong>g reset if <strong>in</strong>let E1 is<br />

open aga<strong>in</strong> <strong>in</strong> standby.<br />

Thermoregulation is <strong>in</strong>active<br />

(Pump off, compressor off,<br />

heat<strong>in</strong>g off).<br />

Advice:<br />

While error message is be<strong>in</strong>g <strong>in</strong>dicated error and set po<strong>in</strong>t are displayed alternately.