Unichiller in Desktop Housing - HUBER
Unichiller in Desktop Housing - HUBER Unichiller in Desktop Housing - HUBER
´L´ Limit – Command This command transfers the set point value limits; Master query: „[M01L0Fllllhhhhpp\r„ llll = Low-Limit, lower set point limit (Format as above) **** = No change of the lower set point limit. hhhh = High-Limit, upper set point limit (Format as above) **** = No change of the upper set point limit. pp = Check sum \r = End-character CR. Slave response: „[S01L17llllhhhhuuuuoooopp\r„ llll = Low-Limit, lower set point limit (Format as above) hhhh = High-Limit, upper set point limit (Format as above) uuuu = Lower working-range limit. This limit is specific to the device and cannot be modified. The lower set point limit cannot be below the lower working range limit. oooo = Upper working range limit. This limit is specific to the device and cannot be modified. The upper set point limit cannot be above the upper working range limit. pp = Check sum \r = End-character CR. 28
Chapter 5: First aid for a fault condition The following sections can be found in this chapter: - Display Error Messages - Maintenance - Decontamination / Repair - Cleaning the surfaces - Plug contacts 29
- Page 1 and 2: Valid for: Operating manual Unichil
- Page 3 and 4: Foreword Dear Customer, The Huber t
- Page 5 and 6: Description of Safety and Informati
- Page 7 and 8: Description Chillers are temperatur
- Page 9 and 10: Safety Devices to DIN12876 - Low le
- Page 11 and 12: Operating conditions Please make su
- Page 13 and 14: Thermofluids We recommend the therm
- Page 15 and 16: Display and operation Display LEDs
- Page 17 and 18: Chapter 3: Connect the machine, fil
- Page 19 and 20: Start up General All models must be
- Page 21 and 22: Setting the over-temperature switch
- Page 23 and 24: Draining the machine and an externa
- Page 25 and 26: Data Communication Transmission for
- Page 27: ´G´ General Command This command
- Page 31 and 32: Maintenance Danger! Prior to carrin
- Page 33 and 34: Chapter 6: Taking the machine out o
- Page 35 and 36: Transport The unit is now decommiss
´L´ Limit – Command<br />
This command transfers the set po<strong>in</strong>t value limits;<br />
Master query: „[M01L0Fllllhhhhpp\r„<br />
llll = Low-Limit, lower set po<strong>in</strong>t limit (Format as above)<br />
**** = No change of the lower set po<strong>in</strong>t limit.<br />
hhhh = High-Limit, upper set po<strong>in</strong>t limit (Format as above)<br />
**** = No change of the upper set po<strong>in</strong>t limit.<br />
pp = Check sum<br />
\r = End-character CR.<br />
Slave response: „[S01L17llllhhhhuuuuoooopp\r„<br />
llll = Low-Limit, lower set po<strong>in</strong>t limit (Format as above)<br />
hhhh = High-Limit, upper set po<strong>in</strong>t limit (Format as above)<br />
uuuu = Lower work<strong>in</strong>g-range limit. This limit is specific to the device and cannot be<br />
modified. The lower set po<strong>in</strong>t limit cannot be below the lower work<strong>in</strong>g range<br />
limit.<br />
oooo = Upper work<strong>in</strong>g range limit. This limit is specific to the device and cannot be<br />
modified. The upper set po<strong>in</strong>t limit cannot be above the upper work<strong>in</strong>g range<br />
limit.<br />
pp = Check sum<br />
\r = End-character CR.<br />