Dra<strong>in</strong><strong>in</strong>g the mach<strong>in</strong>e and an externally closed application<br />

General<br />

- Before dra<strong>in</strong><strong>in</strong>g the unit, the thermal fluid should be at ambient temperature<br />

(approx. 20 °C). If not, let the mach<strong>in</strong>e run with a set-po<strong>in</strong>t of approx. 20 °C for<br />

a few m<strong>in</strong>utes until the thermal fluid is at a safe temperature.<br />

- Connect one end of a suitable dra<strong>in</strong> hose to the dra<strong>in</strong> of the unit (8), and place<br />

the other end <strong>in</strong>to a suitable conta<strong>in</strong>er (make sure the hose and conta<strong>in</strong>er<br />

materials are compatible with the thermal fluid be<strong>in</strong>g used.<br />

Dra<strong>in</strong><strong>in</strong>g<br />

- Open the dra<strong>in</strong> valve (4) if available.<br />

- The thermal fluid should flow through the pump case, and dra<strong>in</strong> port <strong>in</strong>to the<br />

conta<strong>in</strong>er.<br />

- Disconnect the hose from the unit’s circulation flow (1).<br />

- Disconnect the hose from the unit´s circulation return (2).<br />

- Leave the temperature control device open to dry out for some time (without<br />

seal<strong>in</strong>g caps and open dra<strong>in</strong> valve (4).<br />

Chang<strong>in</strong>g thermal fluid / <strong>in</strong>ternal clean<strong>in</strong>g<br />

- After empty<strong>in</strong>g the unit as described <strong>in</strong> the chapter Dra<strong>in</strong><strong>in</strong>g the mach<strong>in</strong>e and an<br />

externally closed application depend<strong>in</strong>g on the thermal fluid, it is possible that<br />

remnants of the oil rema<strong>in</strong> <strong>in</strong> the mach<strong>in</strong>e.<br />

- Connect a short hose between the circulation return (2) and circulation flow (1)<br />

of the unit.<br />

- When hav<strong>in</strong>g used silicon oils as thermal fluid, use a suitable solvent (e.g.<br />

Mucasol) to clean the <strong>in</strong>ternal components such as pump hous<strong>in</strong>g, reservoir, etc.<br />

Depend<strong>in</strong>g on the amount of contam<strong>in</strong>ation, it may be necessary to dra<strong>in</strong> the<br />

solvent off, and repeat the procedure a number of times with clean solvent.<br />

- Afterwards, leave the temperature control device stand for some time (open all<br />

dra<strong>in</strong> valves and have the connections opened).<br />