12.11.2014 Views

Zeiss LSM 510 confocal manual - Duke University Light Microscopy ...

Zeiss LSM 510 confocal manual - Duke University Light Microscopy ...

Zeiss LSM 510 confocal manual - Duke University Light Microscopy ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

1 of 6<br />

<strong>Zeiss</strong> 780 upright <strong>confocal</strong> <strong>manual</strong><br />

Power Up<br />

1. Turn on the X-Cite (if you need to see fluorescence down the eyepieces)<br />

2. Turn on Main Switch<br />

3. Turn on System PC switch<br />

4. Turn on the components switch<br />

5. Turn on the Lasos Argon laser switch (large black box) and turn key to position 1 if you will use it and check it is set<br />

to "Run" not "idle"<br />

6. Turn on the PC<br />

Select the "LMCFuser" account and enter the password<br />

If the system is left on for you check the Argon laser is in run mode.<br />

Open the Software<br />

Double-click the ZEN icon on the desktop and choose "Start System" when the boot status window pops up<br />

Stage configuration<br />

The system is based around an Axio Examiner Z1 upright fixed stage microscope that is designed to have highly<br />

configurable sample holders with a lot of space under the objectives.<br />

The simple stage is usually in place which is good for slides and small samples. The height of this stage can be<br />

crudely adjusted with the objectives raised by loosening the front legs, lowering the condenser and loosening the<br />

thumb screw on the back right side. Now you can carefully move the stage up and down to the appropriate height and<br />

retighten the screw. The line on the stand indicates a good position for slides.<br />

If you need to use a custom stage the simple plate stage and condenser can be entirely removed.<br />

Finding your sample<br />

Choose an objective and lower it to the imaging position. (if you need to swap an objective go to the "Maintain" tab and<br />

select the lens so the system computes correctly spatial measurements and pinhole settings).<br />

Go to the Ocular tab, click online and choose the means of viewing - Brightfield, Blue, Green or Red fluorescence


2 of 6<br />

Focus carefully - the objective moves up and down and the stage is fixed so be careful not to hit your sample with the<br />

objective (not good for either party).<br />

Click "RL Illumination Off" on the touch pad or the shutter off button to turn off fluorescence (if you just use BF leave it<br />

on so the system automatically turns that on each time you go to ocular)<br />

Taking <strong>confocal</strong> images<br />

Click the "Acquisition" tab so you see all the settings for <strong>confocal</strong> imaging. If the "Show <strong>manual</strong> tools" box is<br />

unchecked some of the less common adjustments are hidden (the Show all check marks on each blue bar gives a<br />

similar option).<br />

Setup<br />

If you follow the smart setup path . . .


3 of 6<br />

Manual setup<br />

Or you can <strong>manual</strong>ly change the excitation lasers, emission bands, dichroics etc in the Imaging Setup and <strong>Light</strong> Path<br />

tabs.<br />

The MBS (Main Beam Splitter = dichroic) needs to match the lasers used in the visible and 405 light paths)<br />

One track for simultaneous imaging of the channels, more than one track for sequential scanning<br />

For multi-track: Frame switching should be used if different dichroics, pinholes or overlapping emission bands are<br />

required. Line switching is more convenient and possible when only electronic things change between tracks (eg<br />

AOTF control of excitation, which detector is active)<br />

Ch1 and Ch2 are standard alkali PMTs, ChS1 is an array of 32 high-sensitivity GaAsP PMTs. This can be used as<br />

up to 8 channels in Channel mode (but they have to have the same gain at any time, but different digital gain is<br />

allowed).<br />

If you have created a configuration you can save this for easy reloading in the future<br />

Please ask if you need some help to design an optimal setup for your particular fluorophores and experiment.<br />

Scanning and optimizing your image<br />

Pressing Auto-Exposure is a good starting point to get an image on the screen to optimize. Auto-exposure changes<br />

the detector gain values so something is visible.


4 of 6<br />

Live (this was called FastXY in the <strong>510</strong> software) constantly scans quite quickly at a fairly low resolution to allow you to<br />

make adjustments. (If you need to have the scan parameters you will subsequently adjust to be in effect you can press<br />

continuous).<br />

Adjust the z-position with the coarse/fine focus knob (notice these are on the touch screen pad as well as the<br />

microscope) to choose the plane of interest. If you are taking a z-stack generally choose the brightest plane to adjust<br />

the settings.<br />

Zoom - you can do this with the "Crop" button underneath the image - resize the box that appears - or using the scan<br />

area controls in the Acquisition Mode panel.<br />

Laser powers - adjust the sliders (note the are non-linear) so you have reasonable excitation. The lowest value you<br />

can get away is best to minimize fluorophore saturation and photodamage.<br />

Pinhole - adjust so a suitable optical slice is obtained. A trade-off between z-axis resolution and brightness. A pinhole<br />

of one Airy unit gives essentially gives you the best resolution, opening it slightly from there will allow you to collect<br />

more signal.<br />

Gain/offset optimization - adjust the gain (detector sensitivity) and offset (background level) for each channel.<br />

Selecting range indicator is helpful when doing this. In this mode saturated pixels are highlighted red and 0 intensity<br />

blue, a few red pixels and the background 50% blue is often a good point to aim for.


5 of 6<br />

Averaging and scan speed - Averaging and slower scan rates improve SNR, choose a good balance for quality and<br />

speed.<br />

Sampling rate - The number of pixels in the image can be selected. The optimal button selects the correct number for<br />

a particular objective, wavelength and zoom. This will give you a digital sampling rate adequate to fully sample the<br />

optical resolution.<br />

The acquisition Mode panel also allows you to adjust between unidirectional and bidirectional scanning (faster but<br />

sometimes causes image degradation), and the bit depth of the images (8bit is fine for standard imaging, 12-bit may<br />

help with some quantification).<br />

Press the Snap button to acquire an image with the settings you have specified above.<br />

Saving images<br />

Click the disk icon beneath the open images icons to save the selected image as .lsm files. Generally it is best to save<br />

to the D:/ drive and move the data to a server, USB drive etc after you have closed the software.<br />

The software creates a new tab in most but not all acquisition cases so save as you go to avoid loosing anything.<br />

The <strong>LSM</strong> image format can be open by the free offline viewer or ImageJ (links here). You can export images as TIFF<br />

from the File\Export menu.


6 of 6<br />

Z-stack<br />

Time series<br />

1. Select the Time Series box underneath the Z-series box of above<br />

2. Choose the number of time points and the interval between them<br />

3. Press start experiment<br />

And of course you can do a time series of a stack by having both things active. You will end up with a 4D dataset that<br />

can be saved as a single .<strong>LSM</strong> file.<br />

Ending your session<br />

Retract the objective to the up position and clean any immersion objectives that you used<br />

Move your saved data as appropriate<br />

Sign the logbook<br />

If the next user is scheduled within 2 hr, please put the system to standby<br />

1. Close the Zen2010 software<br />

2. Flick the small switch on Lasos controller board to standby.<br />

If nobody is scheduled to use the <strong>confocal</strong> after you, please power down the system<br />

1. Turn key on Lasos black box (5) to 0 position and wait for about 3 min until fan shuts off.<br />

2. In the meantime, exit Zen2010 software and shut down PC from the Windows Start Menu<br />

3. Turn off System PC and components switches after the cooling fan from 1 is quiet. (3 & 4)<br />

4. Turn off Main Switch (2)<br />

5. Turn off X-Cite (1)<br />

6. Cover the microscope

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!