4. Loosen the bracket thumbscrew and place the<br />

bracket in the track.<br />

5. Very carefully slide the probe through the hole in<br />

the seal tightening screw and advance it until the<br />

tip just protrudes into the chamber. Do not allow<br />

the tip to touch the far side of the well, it could<br />

jam and crack the tip.<br />

6. The final position of the probe tip should have<br />

just the angled portion of the tip exposed with<br />

the bevel facing upward. As the seal is tightened,<br />

it may advance into the chamber. Tighten seal<br />

and then reposition as necessary until the seal no<br />

longer can be moved.<br />

Chamber Plugs<br />

Depending upon the chamber size, different style<br />

plugs may be use, either beveled tip style or O-ring<br />

style with single vertical fill hole.<br />

The beveled type can be rotated to seal off the side<br />

fill hole.<br />

For large volume chambers, e.g. 1000 microliters, the<br />

central hole can be plugged to reduce oxygen leakage<br />

although little reoxygenation actually occurs.<br />

Larger chambers have no side fill hole.<br />

Setting the Optional Beveled Plug Depth<br />

1. Slide a fine tip micro pipette tip into the fill hole<br />

until it touches the far side of the chamber well.<br />

2. Drop the beveled transparent plug into the well<br />

with the bevel matching the angle of the pipette<br />

tip.<br />

3. Place the collar over the plug and tighten the<br />

holding screw facing forward. This will be the<br />

position indicator as well.<br />

4. This process will position the depth of the plug<br />

so that air will clear properly when the chamber<br />

is initially filled<br />

7. Repeat for the second chamber and then attach<br />

the single end of each bifurcated fiber bundle.<br />

Using the Titanium Micro<br />

Chamber with Beveled Plug<br />

1. Fill the chamber with buffer solution that has<br />

been equilibrated at your operating temperature<br />

with the oxygen tension to be used in the<br />

experiment. Use slightly more than the specified<br />

9