Water Jacket Plumbing<br />
1. Use 5/16” ID Tygon tubing on the two outboard<br />
fittings.<br />
2. Secure in place using clamps or cable ties.<br />
3. If the center coupling should require changing,<br />
remove the stirring motor from one block,<br />
remove the block mounting screws and attach<br />
short length of tubing between the inner block<br />
fittings. Again secure tubing. DO NOT<br />
ATTEMPT THIS WITH THE PROBE IN PLACE.<br />
Stirring Motor Installation<br />
1. Loosen the motor holding setscrew (lower front<br />
center of block) until it does not protrude into the<br />
bore.<br />
2. Back off the chamber holding setscrew and<br />
temporarily place the titanium chamber into the<br />
top of the aluminum block.<br />
3. From the bottom, insert the motor/magnet<br />
assembly and push it until it contacts the bottom<br />
of the titanium chamber.<br />
4. Pull the motor back about 1 mm and gently<br />
tighten the motor holding setscrew. Excessive<br />
force will damage the motor.<br />
Titanium Chamber Insertion<br />
and Removal<br />
1. Drop chamber into opening at the top of the<br />
block with the fiber entry hole in the rear and<br />
roughly aligned with the threaded hole that will<br />
accept the fiber sealing screw.<br />
2. Use the fiber sealing screw (without the fiber and<br />
seal) to set the alignment by screwing it into the<br />
block until seated. Some rotation of the chamber<br />
may be necessary.<br />
3. The length of 22 ga.. stainless steel hypodermic<br />
tubing should be inserted prior to installing the<br />
probe in check the alignment once the seal is in<br />
place. Place the seal over the end of the tubing.<br />
4. When the seal screw is seated, tighten the front<br />
chamber holding setscrew. This will assure that<br />
the tip of the fiber will not be damaged when<br />
inserted subsequently. The seal will remain in<br />
the hole once the tubing is removed.<br />
Installing the Probe<br />
Note: At all times take care not to damage the fiber<br />
probe.<br />
1. Attach the probe to the unmounted L-bracket by<br />
screwing it into the ¾” hollow sleeve coupler.<br />
<strong>Fiber</strong> will be on the side nearest the L-bracket<br />
tightening screw. Do not attach the bifurcated<br />
bundle yet. Before tightening the probe, rotate it<br />
until the black angled face is upward to minimize<br />
light leakage.<br />
2. Place the bracket with probe installed on the<br />
groove in the back supporting rail.<br />
3. Loosen the seal tightening screw so that the<br />
probe will pass through the seal without friction.<br />
8