Water Jacket Plumbing<br />

1. Use 5/16” ID Tygon tubing on the two outboard<br />

fittings.<br />

2. Secure in place using clamps or cable ties.<br />

3. If the center coupling should require changing,<br />

remove the stirring motor from one block,<br />

remove the block mounting screws and attach<br />

short length of tubing between the inner block<br />

fittings. Again secure tubing. DO NOT<br />

ATTEMPT THIS WITH THE PROBE IN PLACE.<br />

Stirring Motor Installation<br />

1. Loosen the motor holding setscrew (lower front<br />

center of block) until it does not protrude into the<br />

bore.<br />

2. Back off the chamber holding setscrew and<br />

temporarily place the titanium chamber into the<br />

top of the aluminum block.<br />

3. From the bottom, insert the motor/magnet<br />

assembly and push it until it contacts the bottom<br />

of the titanium chamber.<br />

4. Pull the motor back about 1 mm and gently<br />

tighten the motor holding setscrew. Excessive<br />

force will damage the motor.<br />

Titanium Chamber Insertion<br />

and Removal<br />

1. Drop chamber into opening at the top of the<br />

block with the fiber entry hole in the rear and<br />

roughly aligned with the threaded hole that will<br />

accept the fiber sealing screw.<br />

2. Use the fiber sealing screw (without the fiber and<br />

seal) to set the alignment by screwing it into the<br />

block until seated. Some rotation of the chamber<br />

may be necessary.<br />

3. The length of 22 ga.. stainless steel hypodermic<br />

tubing should be inserted prior to installing the<br />

probe in check the alignment once the seal is in<br />

place. Place the seal over the end of the tubing.<br />

4. When the seal screw is seated, tighten the front<br />

chamber holding setscrew. This will assure that<br />

the tip of the fiber will not be damaged when<br />

inserted subsequently. The seal will remain in<br />

the hole once the tubing is removed.<br />

Installing the Probe<br />

Note: At all times take care not to damage the fiber<br />

probe.<br />

1. Attach the probe to the unmounted L-bracket by<br />

screwing it into the ¾” hollow sleeve coupler.<br />

<strong>Fiber</strong> will be on the side nearest the L-bracket<br />

tightening screw. Do not attach the bifurcated<br />

bundle yet. Before tightening the probe, rotate it<br />

until the black angled face is upward to minimize<br />

light leakage.<br />

2. Place the bracket with probe installed on the<br />

groove in the back supporting rail.<br />

3. Loosen the seal tightening screw so that the<br />

probe will pass through the seal without friction.<br />

8