♦ Analysis Wavelength Box. Enter the<br />

analysis wavelength. The analysis<br />

wavelength should be very close to 600 nm.<br />

(When the excited ruthenium complex at the<br />

tip of the oxygen probe fluoresces, it<br />

typically emits energy at ~600 nm.) In the<br />

Bandwidth [box] pixels area, select the<br />

number of pixels around the analysis<br />

wavelength to average.<br />

♦ Data Display for Sensor Measurement.<br />

Choose the data display format and<br />

precision of the data. Under Format select<br />

Decimal or Scientific. Under Precision,<br />

select a value to specify the precision of the<br />

oxygen data displayed and saved in files.<br />

The maximum is 5.<br />

♦ Wavelength Calibration Coefficients<br />

♦ Do not forget to enter these the first time<br />

using the system.<br />

Check the Wavelength Calibration Data Sheet<br />

that came with your system to make sure the<br />

values on the Data Sheet and in this dialog box<br />

are the same. If not, enter the coefficients<br />

otherwise spectral data may be skewed.<br />

♦ Pressure Compensation CH. If you have<br />

your own external pressure measuring<br />

system, you can use this feature to monitor<br />

and correct for pressure fluctuations in your<br />

sample. You can either use a pressure<br />

transducer separately from the system or<br />

interface it to your sensor system. If you<br />

interface a pressure transducer to your<br />

sensor system, you must have an available<br />

spectrometer channel that is not connected<br />

to an oxygen sensor. Use the pull down<br />

menu to select how you want to monitor<br />

pressure. See Spectrometer DB-25<br />

connector details for attaching the analog<br />

signal to an unused A/D input channel.<br />

♦ Temperature Measurement. If you want to<br />

monitor and correct for temperature<br />

fluctuations in your sample, select a method<br />

from the pull down menu.<br />

-- Select None if you are not monitoring<br />

temperature. For now, this is the<br />

recommended setting.<br />

-- Select Manual if you are monitoring<br />

temperature, but you do not want<br />

OOISensors Software to read and<br />

display the temperature values. The<br />

Manual selection means that you must<br />

manually type temperature values in the<br />

display window. Enable the<br />

Compensate function if you want the<br />

software to correct for temperature<br />

fluctuations only if you calculated the<br />

temperature coefficients. Enabling the<br />

Chart function allows you to view a<br />

chart of the temperature values.<br />

-- Select Omega D5xx1 RS232 if you are<br />

monitoring temperature and if you want<br />

the software to automatically read and<br />

display the temperature values. Ocean<br />

<strong>Optic</strong>s offers the Omega Thermistor and<br />

the Omega Thermocouple for<br />

monitoring temperature. The thermistor<br />

and thermocouple should already be<br />

connected to your PC via an RS-232<br />

module. Next to Serial Port, select the<br />

COM Port number on your PC to which<br />

the thermistor or thermocouple is<br />

connected. Because the RS-232 module<br />

can support up to four thermistors or<br />

thermocouples, it has these four ports<br />

labeled. Next to D5xx1 CH, select the<br />

port to which the thermistor or<br />

thermocouple connects to the RS-232<br />

module. (If you only have one<br />

thermistor or thermocouple, select 0.)<br />

Enable the Compensate function if you<br />

want the software to correct for<br />

temperature fluctuations. Enabling the<br />

Chart function allows you to view a<br />

chart of the temperature values.<br />

♦ Enable reference correction. Disable this<br />

function.<br />

♦ Level alarm. Enable this feature to set<br />

alarm properties. A green indicator appears<br />

in the display window if this feature is<br />

activated. If the values fall below the alarm<br />

parameters, the green indicator turns red. If<br />

the values rise above the alarm parameters,<br />

the green indicator turns yellow.<br />