FLOW CHART NO.6<br />
PANEL+24.5V is not output.<br />
Is approximately +30V voltage supplied to the collector<br />
of Q205?<br />
Check Q202, Q203, D203, P-ON-H1 line and their<br />
periphery circuit, and service it if defective.<br />
FLOW CHART NO.7<br />
PANEL+13V is not output.<br />
Yes<br />
Is approximately +15V voltage supplied to the<br />
collector of Q201?<br />
Is approximately +13V voltage supplied to the<br />
base of Q201?<br />
Replace Q201.<br />
FLOW CHART NO.8<br />
P-ON+7V is not output.<br />
Yes<br />
Yes<br />
Is approximately +7V voltage supplied to the<br />
cathode of D641?<br />
Yes<br />
Check if there is any leak or short-circuit on<br />
the loaded circuit, and service it if defective.<br />
FLOW CHART NO.9<br />
P-ON+5V is not output. (PANEL+13V is outputted normally.)<br />
Is approximately +6.8V voltage supplied to the<br />
collector of Q635?<br />
Is approximately +6V voltage supplied to the<br />
base of Q635 and the base of Q636?<br />
Replace Q635 and Q636.<br />
Yes<br />
Yes<br />
No<br />
No<br />
No<br />
No<br />
No<br />
No<br />
See FLOW CHART No.5<br />
Check C631, D631, D668 and their periphery circuit,<br />
and service it if defective.<br />
Check Q202, Q203, D201, D202, D210 and their<br />
periphery circuit, and service it if defective.<br />
Check C641, D641 and their periphery circuit,<br />
and service it if defective.<br />
See FLOW CHART No.8<br />
Check D659 and their periphery circuit, and service<br />
it if defective.<br />