FLOW CHART NO.6<br />

PANEL+24.5V is not output.<br />

Is approximately +30V voltage supplied to the collector<br />

of Q205?<br />

Check Q202, Q203, D203, P-ON-H1 line and their<br />

periphery circuit, and service it if defective.<br />

FLOW CHART NO.7<br />

PANEL+13V is not output.<br />

Yes<br />

Is approximately +15V voltage supplied to the<br />

collector of Q201?<br />

Is approximately +13V voltage supplied to the<br />

base of Q201?<br />

Replace Q201.<br />

FLOW CHART NO.8<br />

P-ON+7V is not output.<br />

Yes<br />

Yes<br />

Is approximately +7V voltage supplied to the<br />

cathode of D641?<br />

Yes<br />

Check if there is any leak or short-circuit on<br />

the loaded circuit, and service it if defective.<br />

FLOW CHART NO.9<br />

P-ON+5V is not output. (PANEL+13V is outputted normally.)<br />

Is approximately +6.8V voltage supplied to the<br />

collector of Q635?<br />

Is approximately +6V voltage supplied to the<br />

base of Q635 and the base of Q636?<br />

Replace Q635 and Q636.<br />

Yes<br />

Yes<br />

No<br />

No<br />

No<br />

No<br />

No<br />

No<br />

See FLOW CHART No.5<br />

Check C631, D631, D668 and their periphery circuit,<br />

and service it if defective.<br />

Check Q202, Q203, D201, D202, D210 and their<br />

periphery circuit, and service it if defective.<br />

Check C641, D641 and their periphery circuit,<br />

and service it if defective.<br />

See FLOW CHART No.8<br />

Check D659 and their periphery circuit, and service<br />

it if defective.<br />

8-2 A91N3TR