<strong>Test</strong> <strong>procedures</strong> <strong>for</strong> <strong>WDs</strong> <strong>according</strong> <strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-1 (6)<br />

Additions<br />

Source<br />

Requirement<br />

<strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-2 • Definition of dosing accuracy of chemicals of minimum 5 %<br />

Guideline DGKH,<br />

DGSV and AKI <strong>for</strong><br />

Germany<br />

<strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-4<br />

(<strong>WDs</strong> <strong>for</strong> thermolabile<br />

endoscopes)<br />

• Determination to test devices not complying the standards<br />

• Definition reference PCD (Crile-clamps with<br />

blood contamination in the joint)<br />

• Definition of the test load<br />

• Definition of proteine limit value and test of it<br />

• Restatement routine monitoring<br />

• Definition of dosing accuracy of chemicals of minimum 5 %<br />

• Extension and specialization of the scope covering the<br />

requirements <strong>for</strong> endoscope <strong>WDs</strong> (WD-E), e.g.:<br />

• Requirements on the leak tightness of endoscopes<br />

• Requirements on the failure indication of blockage or<br />

missing connection of canals<br />

• Cleaning and flushing of canals<br />

• Chemical disinfection at average temperature<br />

• Self-disinfection of WD-E without endoscopes<br />

using disinfectants or thermal disinfection<br />

5.27-27 546 Dr. U. Kaiser 01/2011