<strong>Test</strong> OQ PQ Rout. Requirement Measurement<br />

Dosage of cleaning<br />

agent<br />

Failure indication<br />

lack of cleaning<br />

agent<br />

<strong>Test</strong> <strong>procedures</strong> <strong>for</strong> <strong>WDs</strong> <strong>according</strong> <strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-1 (4)<br />

Process residuals<br />

Quality of flushing<br />

water<br />

X<br />

X<br />

X<br />

X<br />

X<br />

X<br />

Accuracy of the<br />

dosage <strong>according</strong> to<br />

the manufacturer‘s<br />

instructions<br />

Failure indication,<br />

if the remaining<br />

amount is not<br />

sufficient <strong>for</strong> a cycle<br />

Defined by the<br />

manufacturer of the<br />

chemicals<br />

According to the<br />

instructions of the<br />

WD manufacturer<br />

Dryness of the load X No remaining water<br />

Volumetric or gravimetric<br />

metering<br />

<strong>Test</strong> <strong>for</strong> failure<br />

indication because<br />

of empty reservoir<br />

Ususally by<br />

measurement of pH<br />

and conductivity of<br />

the last flushing<br />

Conductivity, pH,<br />

salts, bioburden<br />

Visual test using<br />

crepe paper or<br />

weight comparison<br />

be<strong>for</strong>e and after the<br />

process<br />

5.25-27 544 Dr. U. Kaiser 01/2011