Validation procedure<br />

of cleaning and disinfection<br />

processes<br />

<strong>according</strong> to <strong>EN</strong> <strong>ISO</strong> <strong>15883</strong> series<br />

5.20-27 538 Dr. U. Kaiser 01/2011


„Accepted guidelines of technology“<br />

<strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-1 to -4<br />

(2009)<br />

Addition:<br />

Guideline of DGKH,<br />

DGSV and AKI <strong>for</strong><br />

Germany (2008)<br />

Requirements and test <strong>procedures</strong> of washer/disinfectors with<br />

• Thermal disinfection <strong>for</strong> surgical instruments, anesthesia<br />

devices, bins, utensils, glass devices etc.<br />

• Thermal disinfection <strong>for</strong> bedpans<br />

• Chemical disinfection <strong>for</strong> thermolabile endoscopes<br />

German recommendation <strong>for</strong> the validation and routine<br />

monitoring of cleaning and disinfection processes <strong>for</strong> medical<br />

devices (MDs) and recommendation <strong>for</strong> the selection of<br />

machines<br />

5.21-27 540 Dr. U. Kaiser 01/2011


<strong>Test</strong> <strong>procedures</strong> <strong>for</strong> <strong>WDs</strong> <strong>according</strong> <strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-1 (1)<br />

1. Type test/test during production<br />

2. Validation during installation with<br />

1. Installation qualification (IQ)<br />

2. Operational Qualification (OQ)<br />

3. Perfomance Qualification (PQ)<br />

3. Routine monitoring (ROUT) and annually requalification (RQ)<br />

5.22-27 541 Dr. U. Kaiser 01/2011


<strong>Test</strong> <strong>procedures</strong> <strong>for</strong> <strong>WDs</strong> <strong>according</strong> <strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-1 (2)<br />

<strong>Test</strong> OQ PQ Rout. Requirement Measurement<br />

Cleaning<br />

<strong>Test</strong> type 1<br />

Cleaning<br />

<strong>Test</strong> type 2<br />

X<br />

X<br />

X<br />

(daily)<br />

Spray system X X X<br />

X<br />

Clean test of the load or<br />

PCD with test soil<br />

<strong>according</strong> <strong>EN</strong> <strong>ISO</strong><br />

<strong>15883</strong>-5 <strong>according</strong> to<br />

country recommendation<br />

Clean test of real<br />

contaminated load<br />

(3 measurements!)<br />

See cleaning<br />

(<strong>Test</strong> type 1)<br />

According<br />

<strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-5<br />

(mostly visual test)<br />

Visual test and<br />

remaining proteine test<br />

• Ninhydrin-Method<br />

• Biuret/BCA-Method<br />

• OPA-Method<br />

See cleaning.<br />

Good: additional<br />

measurement of the<br />

flushing pressure<br />

5.23-27 542 Dr. U. Kaiser 01/2011


<strong>Test</strong> <strong>procedures</strong> <strong>for</strong> <strong>WDs</strong> <strong>according</strong> <strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-1 (3)<br />

<strong>Test</strong> OQ PQ Rout. Requirement Measurement<br />

Temperature profile X X X<br />

Thermal disinfection<br />

reservior<br />

Only <strong>for</strong> chemical<br />

disinfection<br />

X<br />

X<br />

X<br />

Confirmation of<br />

machine settings<br />

and keeping of<br />

tolerances, e.g.:<br />

Cleaning: +/- 5°C<br />

Disinfection -<br />

0/+5°C<br />

Temperature band<br />

≤ 4°C<br />

Confirmation of<br />

machine settings<br />

and achieving<br />

drinking water<br />

quality<br />

Confirmation of<br />

machine settings<br />

(Temperature<br />

profile and dosage)<br />

Temperature<br />

sensor/ data logger<br />

on load carrier,<br />

chamber wall and<br />

load<br />

Temperature<br />

sensor/ data logger/<br />

micro-biological test<br />

Temperature<br />

sensor/ data logger/<br />

Measurement<br />

dosage see below<br />

5.24-27 543 Dr. U. Kaiser 01/2011


<strong>Test</strong> OQ PQ Rout. Requirement Measurement<br />

Dosage of cleaning<br />

agent<br />

Failure indication<br />

lack of cleaning<br />

agent<br />

<strong>Test</strong> <strong>procedures</strong> <strong>for</strong> <strong>WDs</strong> <strong>according</strong> <strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-1 (4)<br />

Process residuals<br />

Quality of flushing<br />

water<br />

X<br />

X<br />

X<br />

X<br />

X<br />

X<br />

Accuracy of the<br />

dosage <strong>according</strong> to<br />

the manufacturer‘s<br />

instructions<br />

Failure indication,<br />

if the remaining<br />

amount is not<br />

sufficient <strong>for</strong> a cycle<br />

Defined by the<br />

manufacturer of the<br />

chemicals<br />

According to the<br />

instructions of the<br />

WD manufacturer<br />

Dryness of the load X No remaining water<br />

Volumetric or gravimetric<br />

metering<br />

<strong>Test</strong> <strong>for</strong> failure<br />

indication because<br />

of empty reservoir<br />

Ususally by<br />

measurement of pH<br />

and conductivity of<br />

the last flushing<br />

Conductivity, pH,<br />

salts, bioburden<br />

Visual test using<br />

crepe paper or<br />

weight comparison<br />

be<strong>for</strong>e and after the<br />

process<br />

5.25-27 544 Dr. U. Kaiser 01/2011


<strong>Test</strong> <strong>procedures</strong> <strong>for</strong> <strong>WDs</strong> <strong>according</strong> <strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-1 (5)<br />

<strong>Test</strong> OQ PQ Rout. Requirement Measurement<br />

Leak tightness of<br />

the chamber<br />

Door locking X X<br />

Reproducibility<br />

Failure indication<br />

X<br />

X<br />

X<br />

No leaks in full<br />

chamber<br />

Not allowed:<br />

• process start with<br />

open door<br />

• Opening the door<br />

during the process<br />

• Opening both<br />

doors<br />

In the last 3 of 4<br />

test cycles the<br />

temperature profile<br />

must be the same<br />

within 2,5°C<br />

Failure indication, if<br />

a sensor is out of<br />

order<br />

Visual test<br />

Check during<br />

operation, if<br />

requirements are<br />

fulfilled<br />

Cooling the WD,<br />

then 4 x recording<br />

temperature profile<br />

with sensors/ data<br />

loggers<br />

Switch off the<br />

sensors separately<br />

and check failure<br />

indication<br />

5.26-27 545 Dr. U. Kaiser 01/2011


<strong>Test</strong> <strong>procedures</strong> <strong>for</strong> <strong>WDs</strong> <strong>according</strong> <strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-1 (6)<br />

Additions<br />

Source<br />

Requirement<br />

<strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-2 • Definition of dosing accuracy of chemicals of minimum 5 %<br />

Guideline DGKH,<br />

DGSV and AKI <strong>for</strong><br />

Germany<br />

<strong>EN</strong> <strong>ISO</strong> <strong>15883</strong>-4<br />

(<strong>WDs</strong> <strong>for</strong> thermolabile<br />

endoscopes)<br />

• Determination to test devices not complying the standards<br />

• Definition reference PCD (Crile-clamps with<br />

blood contamination in the joint)<br />

• Definition of the test load<br />

• Definition of proteine limit value and test of it<br />

• Restatement routine monitoring<br />

• Definition of dosing accuracy of chemicals of minimum 5 %<br />

• Extension and specialization of the scope covering the<br />

requirements <strong>for</strong> endoscope <strong>WDs</strong> (WD-E), e.g.:<br />

• Requirements on the leak tightness of endoscopes<br />

• Requirements on the failure indication of blockage or<br />

missing connection of canals<br />

• Cleaning and flushing of canals<br />

• Chemical disinfection at average temperature<br />

• Self-disinfection of WD-E without endoscopes<br />

using disinfectants or thermal disinfection<br />

5.27-27 546 Dr. U. Kaiser 01/2011