Circulating<br />

water bath<br />

flow<br />

Heat<br />

Exchanger<br />

Heater<br />

Pump<br />

Figure 1. Block diagram of Level 1 ® fluid warmers.<br />

Figure 1 diagrams the H-<strong>1000</strong> and its operation.<br />

An onboard pump circulates warm water through<br />

the aluminum heat exchanger in the disposable<br />

set. Note that the circulating water bath and the<br />

infusate flow against each other, or countercurrent,<br />

which maximizes heat transfer. This<br />

diagram describes the essence of all Level 1 ® fluid<br />

warmers.<br />

System features<br />

Filter with Gas Vent<br />

Level 1 ® D-Series Disposable Sets employ a unique "hands-off" Gas Vent which vents<br />

micro-bubbles of gas that are always released from fluids as they are warmed. In many<br />

other warming systems, these micro-bubbles are actually delivered to the patient along<br />

with the warmed fluids.<br />

The Filter with Gas Vent assembly in the I.V. Sets contains a 170-micron screen filter to<br />

trap macro-aggregates found in blood products. This filter has luer-lock connections and<br />

3