SPECIFICATIONS<br />

• Power Requirements:<br />

• Output<br />

• Output Time Constant<br />

• RMS Noise<br />

• Power-up Transient Period<br />

• Light Source Wavelength<br />

• Sensing Distance (from windows)<br />

• Linearity<br />

7-20 VDC, 3.5 mA avg., 6 mA pk.<br />

0-5.0 VDC<br />

0.1 sec.<br />

< 1 mV<br />

< 1 sec<br />

880 nm<br />

< 5 cm (approx.)<br />

< 2% deviation 0-750 FTU<br />

• Sensitivity/Range Gain Sensitivity (mV/FTU)Range (FTU)<br />

100x 200 25<br />

20x 40 125<br />

5x 10 500<br />

1x 2 **<br />

(** output is non-linear above 750 FTU)<br />

• Temperature Coefficient<br />

• Depth Capability<br />

• Weight (dry)<br />

• Operating Temperature<br />

• Material<br />

• Underwater Connector<br />

< 0.05%/°C<br />

6000 m (19,685 ft)<br />

86 g (3.0 oz)<br />

0°C to 65°C (32°F to 149°F)<br />

ABS plastic, epoxy<br />

Impulse AG-306/206 (others available on request)<br />

DIMENSIONS<br />

Connector Version<br />

Bulkhead Version<br />