CFI60 OBJECTIVES<br />
CFI Plan Fluor Series for Brightfield/Ph<strong>as</strong>e Contr<strong>as</strong>t (DLL)<br />
Featuring an extra-high transmission<br />
rate, especially in the ultraviolet<br />
wavelength, and flatness of field<br />
comparable to the CFI Plan Achromat<br />
series, these objectives are designed<br />
for fluorescence observation and<br />
photomicrography. Because of this<br />
improvement in quality, these<br />
objectives can function <strong>as</strong> multipurpose<br />
objectives for brightfield,<br />
fluorescence, polarizing, and DIC<br />
observations.<br />
The DLL (low contr<strong>as</strong>t) ph<strong>as</strong>e<br />
objectives for this series can also be<br />
used <strong>as</strong> universal objectives for<br />
ph<strong>as</strong>e contr<strong>as</strong>t, fluorescence, DIC,<br />
and brightfield techniques.<br />
5.3<br />
42.8<br />
ø30<br />
M25~0.75<br />
Plan Fluor<br />
4X / 0.13<br />
/ - WD 17.1<br />
ø20<br />
CFI Plan Fluor 4x<br />
45˚<br />
5.3<br />
43.9<br />
ø30<br />
M25~0.75<br />
Plan Fluor<br />
10X / 0.30<br />
DIC L@<br />
/0.17 WD 16.0<br />
ø18<br />
CFI Plan Fluor 10x<br />
CFI Plan Fluor DLL 10x<br />
45˚<br />
1.7<br />
5.3<br />
10.8 11<br />
32<br />
3.9<br />
ø34.5<br />
ø30<br />
M25~0.75<br />
Plan Fluor<br />
60X / 0.85<br />
DIC M<br />
/0.11-0.23 WD 0.30<br />
11 14 17 20 23<br />
ø8.5<br />
ø18<br />
ø22<br />
ø33<br />
30˚<br />
30˚<br />
5.3<br />
3.8 4.5<br />
51.5<br />
ø30<br />
Plan Fluor<br />
100X/1.30 oil<br />
DIC H<br />
/0.17 WD 0.20<br />
ø5.8<br />
ø17<br />
ø21<br />
20˚<br />
45˚<br />
6.2<br />
2.6<br />
5.3<br />
34<br />
2.9 6.5 7.5<br />
17˚<br />
ø33.4<br />
ø32<br />
ø30<br />
M25~0.75<br />
Plan Fluor<br />
100X/0.5-1.3 oil<br />
/0.17 WD 0.2<br />
1.3 0.7<br />
ø5.3<br />
ø19.8<br />
ø23.6<br />
ø32<br />
C2.5<br />
8˚<br />
ø30<br />
M25~0.75<br />
ø30<br />
M25~0.75<br />
ø33.5<br />
ø30<br />
CFI Plan Fluor 60x C<br />
CFI Plan Fluor 100x Oil<br />
CFI Plan Fluor DLL 100x Oil<br />
CFI Plan Fluor 100x Oil with<br />
iris diaphragm<br />
5.3<br />
54<br />
Plan Fluor<br />
20X / 0.50<br />
DIC M<br />
/0.17 WD 2.1<br />
45˚<br />
5.3<br />
57.2<br />
5.5<br />
Plan Fluor<br />
40X / 0.75<br />
DIC M<br />
/0.17 WD 0.72<br />
40<br />
12 5.3<br />
45˚<br />
60.41<br />
Plan Fluor<br />
4X / 0.13<br />
PhL DL<br />
/1.2 WD 16.4<br />
45˚<br />
5<br />
42.8<br />
5.3<br />
M25~0.75<br />
Plan Fluor<br />
40X / 1.30 oil<br />
DIC H<br />
/0.17 WD 0.2<br />
CFI Plan Fluor Series for Ph<strong>as</strong>e Contr<strong>as</strong>t (DL)<br />
These objectives have more contr<strong>as</strong>t than the DLL objectives for enhanced contr<strong>as</strong>t with tissue culture<br />
specimens. These objectives are designed for use on inverted microscopes and therefore their markings are<br />
upside-down for e<strong>as</strong>y reading.<br />
1.2<br />
45˚<br />
3.8<br />
2.7<br />
ø12<br />
ø17<br />
CFI Plan Fluor 20x<br />
CFI Plan Fluor DLL 20x<br />
2<br />
ø10<br />
45˚<br />
CFI Plan Fluor 40x<br />
CFI Plan Fluor DLL 40x<br />
4<br />
3.7<br />
23˚<br />
ø8<br />
ø22<br />
ø25.8<br />
CFI Plan Fluor 40x Oil<br />
60.41<br />
Plan Fluor<br />
10X / 0.30<br />
Ph1 DL<br />
/1.2 WD 15.2<br />
45˚<br />
6<br />
44<br />
5.3<br />
Objectives (magnification) N.A. W.D. (mm) Focal length (mm) Physical depth of focus (µm) Weight (g) Remarks Ph<strong>as</strong>e ring<br />
CFI Plan Fluor 4x 0.13 17.1 50.0 16.27 110<br />
CFI Plan Fluor 10x 0.30 16.0 20.0 3.06 120<br />
CFI Plan Fluor DLL 10x 0.30 16.0 20.0 3.06 120 ph1<br />
CFI Plan Fluor 20x 0.50 2.1 10.0 1.10 175<br />
CFI Plan Fluor DLL 20x 0.50 2.1 10.0 1.10 175 ph1<br />
CFI Plan Fluor 40x 0.75 0.72 5.0 0.49 210 Spring-loaded<br />
CFI Plan Fluor DLL 40x 0.75 0.72 5.0 0.49 210 Spring-loaded ph2<br />
CFI Plan Fluor 40x Oil 1.30 0.2 5.0 0.25 230 Spring-loaded Stopper<br />
CFI Plan Fluor 60x C 0.85 0.3 3.3 0.38 265 Spring-loaded Cover gl<strong>as</strong>s correction 0.11–0.23mm<br />
CFI Plan Fluor 100x Oil 1.30 0.2 2.0 0.25 250 Spring-loaded Stopper<br />
CFI Plan Fluor DLL 100x Oil 1.30 0.2 2.0 0.25 250 Spring-loaded Stopper ph3<br />
CFI Plan Fluor 100x Oil with iris diaphragm 0.5–1.3 0.2 2.0 0.25 285 Spring-loaded With iris diaphragm<br />
M25~0.75<br />
M25~0.75<br />
ø30<br />
ø30<br />
CFI Plan Fluor DL 4x<br />
CFI Plan Fluor DL 10x<br />
Objectives (magnification) N.A. W.D. (mm) Focal length (mm) Physical depth of focus (µm) Weight (g) Remarks Ph<strong>as</strong>e ring<br />
CFI Plan Fluor DL 4x 0.13 16.4 50.0 16.27 105 Cover gl<strong>as</strong>s thickness 1.2mm phL<br />
CFI Plan Fluor DL 10x 0.30 15.2 20.0 3.06 120 Cover gl<strong>as</strong>s thickness 1.2mm ph1<br />
8<br />