Calibrating the turbine<br />

<br />

Note: if you are using a slow<br />

PC, we recommend to set<br />

an higher refresh time.<br />

After having run the calibration program:<br />

1. Select Reference Values from the File menu. If your syringe has a different value<br />

from the default one (3 litres), please enter the correct value.<br />

2. Select Calibration/Turbine….<br />

3. When the Calibration Turbine dialog box appears with the syringe piston initially<br />

pushed all the way in, move the piston in and out for 5 inspiratory strokes and 5<br />

expiratory strokes in order to get the first values appearing on the display. Then<br />

move the syringe piston for other 10 strokes (IN and EX).<br />

4. At each of the 10 steps the software displays the results of the manoeuvre and the<br />

percentage error in the reading.<br />

5. At the end of this operation, the software displays the new calibration factors. Press<br />

OK to store the new value.<br />

48 - <strong>Quark</strong> <strong>PFT</strong> <strong>User</strong> Manual