03.11.2012 Views

Weiss GWE safety workbenches for laboratories

Weiss GWE safety workbenches for laboratories

Weiss GWE safety workbenches for laboratories

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Weiss</strong> <strong>GWE</strong> <strong>safety</strong> <strong>workbenches</strong><br />

<strong>for</strong> <strong>laboratories</strong>


2<br />

<strong>Weiss</strong> <strong>GWE</strong><br />

Effectively against contamination<br />

<strong>Weiss</strong> <strong>GWE</strong> HFC-SH 900<br />

<strong>Weiss</strong> <strong>GWE</strong> <strong>safety</strong> workbench. Increased application<br />

in the chemical-pharmaceutical industry<br />

Nowadays, a sensible approach to<br />

dealing with active and unknown<br />

substances in <strong>laboratories</strong> within<br />

the chemical-pharmaceutical in-<br />

dustry calls <strong>for</strong> safe workstations,<br />

in particular with regard to weighing<br />

work.<br />

This is guaranteed particularly by<br />

the workbench HFC-SH 900.<br />

In the HFC-S <strong>safety</strong> workbench,<br />

personnel and the surrounding<br />

environment are protected effectively<br />

against contamination by the<br />

systematic use of filters and an<br />

optimized air flow.<br />

The bright and transparent working<br />

area and ergonomic design provide<br />

a com<strong>for</strong>table working environment.<br />

Application<br />

• Work with active and toxic<br />

powdered substances (active<br />

ingredients)<br />

• Weighing work (micro and<br />

semi-micro balances according<br />

to USP requirements)<br />

• Sampling<br />

• Product transfer<br />

• Transfer works


Perfect Protection <strong>for</strong> person, product and environment<br />

Maximum <strong>safety</strong><br />

• Protection from airborne<br />

particles and aerosols by the<br />

inward air flow<br />

• Optimized air flow design<br />

ensures the highest possible<br />

level of <strong>safety</strong> and Robustness<br />

• Increased <strong>safety</strong> due to two<br />

consecutive filters (HEPA H14)<br />

In order to ensure the protection<br />

of the user over several hours of<br />

operation, the increasing pressure<br />

drop across the filters is adjusted<br />

automatically<br />

The innovative bag-out filter system<br />

facilitates low-contamination replace-<br />

ment of the pre-filter.<br />

Main filter can be qualified<br />

• The main filter can be qualified<br />

in situ by means of DEHS filter<br />

scanning contaminated air is fed<br />

exclusively through the particle<br />

filter<br />

• The positive pressure plenum<br />

ensures that the pre-filter is<br />

dynamically sealed and the<br />

contaminated air is fed exclu-<br />

sively through the particle filter<br />

• The innovative bag-out filter<br />

change is carried out with low<br />

contamination risk within the<br />

work chamber during operation<br />

Maximum quietness -<br />

only 52 dB(A)<br />

Optimal air flow and sound-ab-<br />

sorbing measures have enabled<br />

us to reduce the noise level to a<br />

minimum.<br />

HEPA H14 filter<br />

Operating Mode<br />

Exhaust air<br />

• In this containment, the air flows<br />

inwards through the filter.<br />

• The air flow is guided towards<br />

the interior in order to prevent<br />

released particles and aerosols<br />

from leaving the controlled<br />

working area.<br />

• The two consecutive HEPA H14<br />

filters retain the particles effectively.<br />

• The optimized air flow design<br />

ensures the highest possible level<br />

of <strong>safety</strong> and robustness.<br />

Airflow and filtering in the HFC-SH 900<br />

Illumination<br />

Tested and certified<br />

Intake air<br />

Tested personal protection in<br />

accordance with EN 14175-3 <strong>for</strong><br />

laboratory fume cupboards. Escape<br />

of < 10 ìg/m 3 breathing air measured<br />

in an onsite test.<br />

3


4<br />

<strong>Weiss</strong> <strong>GWE</strong><br />

Equipments and options <strong>for</strong> your <strong>safety</strong> <strong>workbenches</strong><br />

The workbench can be adapted to<br />

meet your additional requirements:<br />

Active carbon cassette with pre-filter<br />

Anti-glare lighting <strong>for</strong> good visibility<br />

Underframe <strong>for</strong> maximum legroom and<br />

stability<br />

*only in combination with underframe<br />

Waste disposal system <strong>for</strong> safe handling of<br />

waste and samples<br />

Ionization unit <strong>for</strong> electrostatic discharge of<br />

powders and equipment, placed underneath<br />

the arm support and effective across the<br />

whole work surface<br />

Robust work surface made of glazed technical<br />

ceramic and is chemically resistant, highly<br />

scratch-resistant, shock-proof and easy to<br />

clean<br />

• Exhaust air connection connection:<br />

to the building exhaust air<br />

system<br />

• Cable pass-through sealable side<br />

openings <strong>for</strong> supply and data<br />

lines<br />

• Spillage protection from lipped<br />

edges<br />

• Excellent visibility from an er-<br />

gonomic seating position thanks<br />

to sloping front cover<br />

• Hinged front cover <strong>for</strong> easier<br />

insertion of larger objects<br />

• Solvent sensor to detect satura-<br />

tion of active carbon


Perfect Protection <strong>for</strong> person, product and environment<br />

Precise weighing and protection of the operator<br />

<strong>Weiss</strong> <strong>GWE</strong> HFC-SH 1200 MT<br />

The HFC-SH 1200 MT combines<br />

precise weighing technology and<br />

protection of the operator while<br />

handling active and toxic sub-<br />

stances. In addition to the features<br />

of the HFC-S series, the MT version<br />

offers the following advantages <strong>for</strong><br />

high resolution scales:<br />

Low vibration<br />

The scale table and working area<br />

are completely decoupled by a<br />

table-in-table solution. This prevents<br />

the transfer of vibrations and<br />

facilitates work with high-resolution<br />

scales as per USP requirements.<br />

Free work surface<br />

In order to conceal the cabling,<br />

it is fed out underneath the scale<br />

through the centre of the table.<br />

Supply and data lines can be fed<br />

out through the sides via additional<br />

openings.<br />

Example of a minimum weight according<br />

to USP<br />

Minimum weights<br />

The determination of minimum<br />

weights according to USP must be<br />

carried out onsite with the actual<br />

measurement setup and <strong>for</strong>ms a<br />

part of the per<strong>for</strong>mance qualifi-<br />

cation (PQ) process.<br />

For this <strong>safety</strong> workbench generic<br />

minimum weights are determined<br />

which can be used as empirical<br />

values in the design qualification<br />

process.<br />

5


6<br />

<strong>Weiss</strong> <strong>GWE</strong><br />

<strong>Weiss</strong> <strong>GWE</strong> HFC-S 1300<br />

with increased interior<br />

Comparison HFC-SH 900 with HFC-S 1300. On the right: The <strong>safety</strong> workbench<br />

with larger interior is perfect <strong>for</strong> greater space requirement.<br />

A larger interior ensures even<br />

safer working conditions at the<br />

HFC-S 1300.<br />

• All features of the HFC-S series<br />

have been implemented in the<br />

1300 version, together with a<br />

considerably larger interior.<br />

• The working height of the interior<br />

has been increased from 500 mm<br />

to approx. 900 mm. Devices with<br />

a total height of approx. 830 mm<br />

can easily be inserted through the<br />

enlarged front opening.<br />

• Typical usage includes decanting,<br />

sampling, weighing large quan-<br />

tities of substances and working<br />

with a sifting machine.<br />

• Regulation and control system<br />

SKANCONTROL SKC4. Control of<br />

the different functions of the<br />

<strong>safety</strong> <strong>workbenches</strong>. Simple operation<br />

with indicators and pushbuttons.


Perfect Protection <strong>for</strong> person, product and environment<br />

<strong>Weiss</strong> <strong>GWE</strong> HFC-SH 1200 Patho<br />

Safety in pathology <strong>laboratories</strong><br />

This <strong>safety</strong> workbench protects you<br />

reliably in your work in pathology<br />

and histology <strong>laboratories</strong> against<br />

inhaling solvent vapours. The table<br />

extraction prevents the release of<br />

vapours, e.g. <strong>for</strong>maldehyde or<br />

toluene<br />

• Solvents, and especially <strong>for</strong>maldehyde<br />

vapours, are effectively<br />

removed from the exhaust air via<br />

an active carbon filter<br />

• Convenient cleaning of working<br />

area, which is made of stainless<br />

steel and glass<br />

• The open construction ensures<br />

ergonomic working conditions in<br />

a well-lit working area<br />

Air flow through a HFC-SH 1200 Patho<br />

Active carbon filter<br />

HEPA H14 filter<br />

Illumination<br />

Lateral<br />

pass-through<br />

Operating Mode<br />

• The air flow, which is directed downwards and towards the<br />

rear (green arrows), guides the released solvent vapours<br />

away from the user. This protective air flow is produced by<br />

a lownoise ventilator.<br />

• HEPA H14 filters are placed both in front of and behind the<br />

ventilator in order to remove aerosols and particles from the<br />

exhaust air.<br />

• The solvent vapours that are removed from the air are<br />

collected in the active carbon filter cassettes.<br />

• The cassettes contain a special type of active carbon that<br />

adsorbs all common solvents and has been prepared <strong>for</strong> use<br />

with <strong>for</strong>maldehyde.<br />

• The <strong>safety</strong> workbench has a pleasant noise level (approx.<br />

52 d(B)A). The ventilator is adjusted automatically if filter<br />

resistance increases (filter blockage).<br />

7


Technical data<br />

Series HFC-SH 900 HFC-SH 1200 HFC-S 1300 HFC-SH 1200 Patho<br />

Outer dimensions<br />

(W × D × H)<br />

(H with active carbon kit)*<br />

(H with draught diverter)<br />

Underframe<br />

(W × D × H)<br />

Inner dimensions working area<br />

<strong>Weiss</strong> <strong>GWE</strong> GmbH<br />

Wiechmannsalle 3<br />

27798 Hude<br />

Germany<br />

Phone: +49 (0) 4484 /189-0<br />

Fax: +49 (0) 4484 /189-189<br />

contact@gwe.de<br />

www.gwe.de<br />

mm<br />

mm<br />

mm<br />

mm<br />

mm<br />

(W × D × H) mm<br />

Height of work opening<br />

with: closed front cover<br />

open front cover<br />

Separation efficiency of<br />

HEPA H14<br />

Intake/exhaust air<br />

volumetric flow rate<br />

Stand-by mode<br />

in operation<br />

mm<br />

mm<br />

mm<br />

%<br />

µm<br />

m 3 /h<br />

m 3 /h<br />

900 x 740 x 1075<br />

H + 160<br />

H + 110<br />

850 x 710 x 760<br />

Adjustability H: + 40/-10<br />

835 x 545 x 600<br />

270<br />

470<br />

99,995<br />

0,3<br />

160<br />

280<br />

1200 x 740 x 1075<br />

H + 160<br />

H + 110<br />

1150 x 710 x 760<br />

Adjustability H: + 40/-10<br />

1135 x 545 x 600<br />

270<br />

470<br />

99,995<br />

0,3<br />

1215 x 880 x 1460<br />

H + 355<br />

H + 110<br />

1150 x 850 x 760<br />

Adjustability H: + 40/-10<br />

1136 x 655 x 900<br />

1210 x 720 x 1236**<br />

1210 x 940 x 1236***<br />

inkl.<br />

H + 110<br />

1150 x 710 x 760<br />

Adjustability H: + 40/-10<br />

1130 x 440 x 570**<br />

1130 x 660 x 570***<br />

545<br />

835 498<br />

99,995<br />

Weight incl. underframe kg 160 200 250 150<br />

220<br />

380<br />

0,3<br />

445<br />

780<br />

99,995<br />

Electrical connection V/Hz 230/50 230/50 230/50 230/50<br />

Power consumption approx. W 195 250 400 400<br />

Light intensity Lux (lx) > 700 > 700 > 700 > 700<br />

Noise level dB (A) 52 52 55 52<br />

* Required free height above the outlet: > 90 mm, ** Type: HFC-SH 1200 Patho / 700, *** Typ: HFC-SH 1200 Patho / 900<br />