03.11.2014 Views

INSTRUmENTS - Dentsply

INSTRUmENTS - Dentsply

INSTRUmENTS - Dentsply

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Sterilisation for<br />

Hand Instruments,<br />

Burs and Cavitron<br />

HAND INSTRUMENTS<br />

WARNINGS<br />

Instruments are not sterile upon receipt and must be sterilised before<br />

use in accordance with the following instructions. Instruments can be<br />

damaged by alkaline and acidic detergents. Always use a pH neutral<br />

detergent or solution. Do not exceed 140°C. Remove all packaging<br />

prior to cleaning and sterilisation. Contaminated instruments should<br />

be handled with gloves and eye protection should be used.<br />

LIMITATIONS ON REPROCESSING: Repeated processing has<br />

minimal effect on these instruments. End of life is normally determined<br />

by wear and damage due to use. See Inspection. Never clean<br />

colour-etched probes in ultrasonic baths. Do not subject Flexichange<br />

instruments to dry heat sterilisation.<br />

INSTRUCTIONS<br />

POINT OF USE: Remove all excess soil with a disposable cloth<br />

paper wipe.<br />

CONTAINMENT AND TRANSPORTATION: Protect delicate or fine<br />

instruments from damage. Instruments should be reprocessed as soon<br />

as is reasonably practical following use. Instruments left standing wet<br />

may stain or corrode.<br />

PREPARATION FOR CLEANING: No specific requirements.<br />

Disassembly not required.<br />

AUTOMATIC CLEANING: Use only validated washing machines<br />

(HTM 0101, HTM 0105: UK only) and validated cleaning agents<br />

suitable for both instruments and washing machine. Follow instructions<br />

issued by both the equipment and cleaning agent manufacturers.<br />

When unloading, inspect for complete removal of visible soil. If<br />

necessary, repeat cycle, or use manual cleaning before<br />

repeating cycle.<br />

MAINTENANCE: Apply a small quantity of surgical grade lubrication<br />

oil to forceps hinges prior to sterilisation.<br />

INSPECTION AND FUNCTION TESTING: Visually inspect to<br />

ensure all contamination has been removed. Check for distortion,<br />

damage and wear. Cutting edges should be free of defects. Discard<br />

damaged, worn or corroded instruments. Ensure that detachable tips<br />

are secure.<br />

DRYING: Dry using paper towelling or dry heat not exceeding<br />

140°C.<br />

PACKAGING FOR STERILISATION: The UK Department of Health<br />

recommends that validated, pre-vacuum autoclaves are used for<br />

sterilising wrapped instruments. Autoclaves drawing only a poststerilisation<br />

vacuum for drying purposes are not suitable for processing<br />

wrapped instruments. Instruments may be loaded into dedicated<br />

instrument trays or general- purpose sterilisation trays. Ensure that<br />

cutting edges are protected.<br />

STERILISATION: Use only validated vacuum autoclaves (Refer to HTM<br />

0101, HTM 0105: UK only). Do not exceed 140°C.<br />

Sterilisation Temperature Range (°C) Nominal Pressure (MPa) Minimum Time at Temperature* (mins)<br />

Minimum<br />

Maximum<br />

134 137 0.225 3<br />

* Minimum holding time when nominal operating temperature and pressure have been reached.<br />

STORAGE: Avoid contamination.<br />

ADDITIONAL INFORMATION: When sterilising multiple instruments<br />

in one autoclave cycle, ensure that the manufacturers stated maximum<br />

load is not exceeded. Validated ultrasonic baths may be used with<br />

validated cleaning agents. Follow instructions for use issued by both<br />

the equipment and cleaning agent manufacturers. MANUFACTURER<br />

CONTACT: See brochure for telephone and address of local<br />

representative or telephone DENTSPLY Ash Instruments on<br />

+44 (0)1932 853422.<br />

DIAMOND/TUNGSTEN CARBIDE<br />

DENTAL BURS: STERILISATION INSTRUCTIONS<br />

Manufacturer: DENTSPLY Limited.<br />

WARNINGS<br />

Burs are not sterile upon receipt and must be sterilised before use in<br />

accordance with the following instructions. Burs can be damaged by<br />

alkaline and acidic detergents. Always use a pH neutral detergent<br />

or solution. Do not exceed 140°C. Remove all packaging prior to<br />

cleaning and sterilisation. Contaminated burs should be handled<br />

with gloves and eye protection should be used.<br />

LIMITATIONS ON REPROCESSING: Repeated processing has<br />

minimal effect on these burs. End of life is normally determined by<br />

wear and damage due to use. See Inspection. Note: Steel Burs are<br />

single use only. Cannot be reprocessed. See Inspection.<br />

INSTRUCTIONS<br />

POINT OF USE: Remove from packaging. Diamond/tungsten burs:<br />

Remove all excess soil with disposable cloth paper wipe.<br />

CONTAINMENT AND TRANSPORTATION: Protect delicate or fine<br />

206

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!