Local Anaesthetic Armamentarium<br />

This document clarifies the correct syringe to be used with the two different types of DENTSPLY<br />

cartridges available to allow aspiration, which is considered the best standard of care.<br />

Cartridge Syringe Piston Style Result Recommended<br />

Self-aspirating<br />

cartridge (a hole<br />

in the plunger/bung)<br />

Cylindrical Passive/auto aspiration YES<br />

Conical Manual aspiration YES<br />

Flat ended Aspiration not permitted therefore not recommended NO<br />

Barbed<br />

Not recommended – cartridge will leak and aspiration<br />

not possible<br />

NO<br />

Standard cartridge<br />

(a solid plunger/bung)<br />

Cylindrical Not recommended NO<br />

Conical Aspiration not permitted therefore not recommended NO<br />

Flat ended Aspiration not permitted therefore not recommended NO<br />

Barbed Manual aspiration YES<br />

DENTSPLY Syringe that can be used with STANDARD cartridges<br />

The barbed piston pierces the solid plunger/bung of a standard cartridge allowing manual aspiration.<br />

DENTSPLY product code: 62413105<br />

DENTSPLY product code: 62413105<br />

DENTSPLY Syringe product that code: can 62413105 be used with SELF-ASPIRATING cartridges<br />

This syringe has a cylindrical piston that fits in the self-aspirating cartridge to allow passive/auto aspiration.<br />

Also known as Astra style.<br />

DENTSPLY product code: 62413200<br />

DENTSPLY product code: 62413200<br />

Mechanism of the DENTSPLY Self-Aspirating Cartridge<br />

The advantage of the passive aspiration system is that no active movement of the operator’s fingers or thumb is required<br />

to effect DENTSPLY aspiration, product and code: thus the 62413200 chance of any unwanted movement of the needle tip is reduced.<br />

• Self-aspiration cartridges have a specially adapted bung, which<br />

needs a different syringe plunger.<br />

DENTSPLY product code: 62413105<br />

• The thin centre of the bung is distended when thumb<br />

pressure is exerted.<br />

• When force is stopped the bung returns to usual shape causing<br />

negative pressure (aspiration).<br />

DENTSPLY product code: 62413200<br />