ELECTROENCEPHALOGRAPHY LABORATORY (<strong>EEG</strong>) NOVEMBER 29, 2011<br />

<strong>Design</strong> Conditions<br />

Year-round Conditions: 70°F to 75°F [21°C to 24°C] and 20% to 60% RH as defined in 2008 ASHRAE<br />

Standard 170 (including amendment d). The system shall be capable <strong>of</strong> maintaining temperatures within the<br />

range during normal working conditions. The cooling load for these spaces shall be calculated to maintain 75°F<br />

[24°C] at 60% RH and the heating load shall be calculated to maintain 70°F [21°C] at 20% RH. The year-round<br />

conditions can be used for variable air volume (VAV) or constant volume (CV) systems. Year-round design<br />

conditions shall be used for all patient areas. See March 2011 VA HVAC <strong>Design</strong> Manual for further details.<br />

Plumbing<br />

General<br />

Plumbing Systems shall be designed as required to meet the needs <strong>of</strong> the <strong>EEG</strong> Lab and <strong>EEG</strong>/EMG individual<br />

rooms. Follow criteria in VA Technical Information Library (TIL) Plumbing <strong>Design</strong> Manual (PG-18-10) listed<br />

on VA website under <strong>Office</strong> <strong>of</strong> Construction & Facilities Management (CFM). Also Refer to (PG-18-1) Master<br />

Construction Specifications and (PG-18-4) Standard Details and CAD Standards for items that may apply<br />

within the <strong>EEG</strong>.<br />

Room Data Sheet Criteria<br />

The Room Data Sheets in Section Four indicate typical quantities <strong>of</strong> plumbing fixtures and equipment as well<br />

as medical gas outlets. They are provided to establish the general planning parameters. The design architect<br />

and engineer (A/E) shall verify the exact fixtures and medical gas locations and quantities for individual<br />

projects.<br />

Water Systems<br />

Domestic cold and hot water shall be piped to all plumbing fixtures and equipment requiring these utilities. A<br />

hot water return system shall be provided to ensure the design temperature is met at the fixture furthest from<br />

the source.<br />

Waste Water Systems<br />

Plumbing fixtures and drains shall be drained by gravity through sanitary waste stacks, including required vent<br />

stacks. Fixtures located below gravity drain line shall be pumped as required by a duplex ejector system. Any<br />

special acidic waste shall be drained through corrosion-resistant, flame-retardant piping into either a local or<br />

centralized acid dilution tank.<br />

Medical Gas Systems<br />

If medical gases are required, their distribution is noted on guide plates for key selected rooms. The design A/E<br />

shall refer to the latest edition (2010) <strong>of</strong> the Facility <strong>Guide</strong>lines Institute (FGI) <strong>Guide</strong>lines for the <strong>Design</strong> and<br />

Construction <strong>of</strong> Healthcare Facilities for minimum quantities and locations <strong>of</strong> medical gases. Specific quantities<br />

and locations should be determined on a per project basis.<br />

FGI notes minimum gas requirements. Specific quantities and locations should be planned with facility user<br />

advisor groups on a per project basis<br />

DESIGN STANDARDS<br />

SECTION 4 - PAGE 38