datasheet - Dolomite Microfluidics
datasheet - Dolomite Microfluidics
datasheet - Dolomite Microfluidics
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
The <strong>Dolomite</strong> Centre Ltd<br />
Benefits<br />
Allows the flow rate range of the Mitos P-Pump to be maximized<br />
Enables the user to operate at very low flow rates with the Mitos P-Pump<br />
Enables the use of very low viscosity liquids (e.g. gases) with the Mitos P-Pump<br />
Prevents flow rate instability when using fluids of different viscosities<br />
Two Flow Resistors can be used to provide equal flow splitting into two channels<br />
Quick and easy to connect using included fittings<br />
Excellent chemical compatibility<br />
Specifications<br />
Specification 3200265<br />
F0.1<br />
Approx. flow<br />
rate generated<br />
by water<br />
pumped at<br />
1bar pressure<br />
3200266<br />
F0.3<br />
3200267<br />
F1<br />
3200268<br />
F3<br />
3200269<br />
F10<br />
3200270<br />
F30<br />
3200271<br />
F100<br />
0.1µl/min 0.3µl/min 1µl/min 3µl/min 10µl/min 30µl/min 100µl/<br />
min<br />
Max flow rate 1 1µl/min 3µl/min 10µl/min 30µl/min 100µl/<br />
min<br />
300µl/<br />
min<br />
1000µl/<br />
min<br />
Min flow rate 2<br />
Internal<br />
diameter<br />
Internal<br />
diameter<br />
tolerance<br />
Outer diameter<br />
Wetted<br />
materials<br />
Sheath<br />
materials<br />
0.002µl/<br />
min<br />
0.006µl/<br />
min<br />
0.02µl/mi<br />
n<br />
0.06µl/mi<br />
n<br />
0.2µl/min 0.6µl/min 2µl/min<br />
25µm 50µm 75µm 100µm<br />
± 1µm ± 2µm ± 3µm ± 3µm<br />
1/16” (1.59mm)<br />
Fused silica<br />
Polyimide (inner sheath)<br />
PEEK (outer sheath)<br />
1) Maximum flow rate is the approximate flow rate produced with water when the Mitos P-<br />
Pump is set to 10bar and there is little or no fluidic resistance downstream of the flow<br />
restrictor.<br />
2) Minimum flow rate is the approximate flow rate produced with water when the Mitos P-<br />
Pump is set to 20mbar and there is little or no fluidic resistance downstream of the flow<br />
restrictor.<br />
MAR-000029 1.5 Page 4 of 22