Why is the baseline noisy, the migration of my sample slow, and the resolution<br />

of the variants poor?<br />

If the <strong>HT</strong> <strong>Protein</strong> Express, the <strong>HT</strong> Low MW <strong>Protein</strong>, the <strong>HT</strong> Pico<br />

<strong>Protein</strong>, or the ProfilerPro Glycan Profiling assay was run on the<br />

instrument, then the electrodes must be cleaned before running<br />

the <strong>HT</strong> <strong>Protein</strong> <strong>Charge</strong> <strong>Variant</strong> assay. If the electrodes are not<br />

cleaned, then residual gel will contaminate the chip and<br />

adversely affect the performance. To fix this problem, remove the<br />

chip from the instrument, aspirate and rinse the wells of the chip<br />

as described in the Chip Preparation section, and add fresh<br />

Running Buffer. Next, replace the buffer tube with a new tube<br />

containing fresh Running Buffer. Then, thoroughly clean the<br />

electrodes of the instrument, using molecular biology-grade<br />

water and a swab. Place the chip back in the instrument and<br />

restart the run.<br />

_________________________________________________________________________________________<br />

Caliper - a <strong>PerkinElmer</strong> Company Page 14 of 18 PN: CLS135474 Rev. 02<br />

68 Elm Street<br />

Hopkinton, MA 01748-1668<br />

1-877-LABCHIP (1-877-522-2447)

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!