Why is the baseline noisy, the migration of my sample slow, and the resolution<br />
of the variants poor?<br />
If the <strong>HT</strong> <strong>Protein</strong> Express, the <strong>HT</strong> Low MW <strong>Protein</strong>, the <strong>HT</strong> Pico<br />
<strong>Protein</strong>, or the ProfilerPro Glycan Profiling assay was run on the<br />
instrument, then the electrodes must be cleaned before running<br />
the <strong>HT</strong> <strong>Protein</strong> <strong>Charge</strong> <strong>Variant</strong> assay. If the electrodes are not<br />
cleaned, then residual gel will contaminate the chip and<br />
adversely affect the performance. To fix this problem, remove the<br />
chip from the instrument, aspirate and rinse the wells of the chip<br />
as described in the Chip Preparation section, and add fresh<br />
Running Buffer. Next, replace the buffer tube with a new tube<br />
containing fresh Running Buffer. Then, thoroughly clean the<br />
electrodes of the instrument, using molecular biology-grade<br />
water and a swab. Place the chip back in the instrument and<br />
restart the run.<br />
_________________________________________________________________________________________<br />
Caliper - a <strong>PerkinElmer</strong> Company Page 14 of 18 PN: CLS135474 Rev. 02<br />