21.10.2014 Views

European Resuscitation Council Guidelines for Resuscitation 2010 ...

European Resuscitation Council Guidelines for Resuscitation 2010 ...

European Resuscitation Council Guidelines for Resuscitation 2010 ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

1318 C.D. Deakin et al. / <strong>Resuscitation</strong> 81 (<strong>2010</strong>) 1305–1352<br />

The tubes are sized in millimetres according to their internal<br />

diameter, and the length increases with diameter. The traditional<br />

methods of sizing a nasopharyngeal airway (measurement against<br />

the patient’s little finger or anterior nares) do not correlate with the<br />

airway anatomy and are unreliable. 326 Sizes of 6–7 mm are suitable<br />

<strong>for</strong> adults. Insertion can cause damage to the mucosal lining of the<br />

nasal airway, resulting in bleeding in up to 30% of cases. 327 If the<br />

tube is too long it may stimulate the laryngeal or glossopharyngeal<br />

reflexes to produce laryngospasm or vomiting.<br />

Oxygen<br />

During CPR, give oxygen whenever it is available. There are no<br />

data to indicate the optimal arterial blood oxygen saturation (SaO 2 )<br />

during CPR. There are animal data 328 and some observational clinical<br />

data indicating an association between high SaO 2 after ROSC and<br />

worse outcome. 329 A standard oxygen mask will deliver up to 50%<br />

oxygen concentration, providing the flow of oxygen is high enough.<br />

A mask with a reservoir bag (non-rebreathing mask), can deliver<br />

an inspired oxygen concentration of 85% at flows of 10–15 l min −1 .<br />

Initially, give the highest possible oxygen concentration. As soon<br />

as the arterial blood oxygen saturation can be measured reliably,<br />

by pulse oximeter (SpO 2 ) or arterial blood gas analysis, titrate the<br />

inspired oxygen concentration to achieve an arterial blood oxygen<br />

saturation in the range of 94–98%.<br />

Suction<br />

Use a wide-bore rigid sucker (Yankauer) to remove liquid (blood,<br />

saliva and gastric contents) from the upper airway. Use the sucker<br />

cautiously if the patient has an intact gag reflex; pharyngeal stimulation<br />

can provoke vomiting.<br />

Ventilation<br />

Provide artificial ventilation as soon as possible <strong>for</strong> any patient in<br />

whom spontaneous ventilation is inadequate or absent. Expired air<br />

ventilation (rescue breathing) is effective, but the rescuer’s expired<br />

oxygen concentration is only 16–17%, so it must be replaced as soon<br />

as possible by ventilation with oxygen-enriched air. The pocket<br />

resuscitation mask is used widely. It is similar to an anaesthetic<br />

facemask, and enables mouth-to-mask ventilation. It has a unidirectional<br />

valve, which directs the patient’s expired air away from<br />

the rescuer. The mask is transparent so that vomit or blood from the<br />

patient can be seen. Some masks have a connector <strong>for</strong> the addition<br />

of oxygen. When using masks without a connector, supplemental<br />

oxygen can be given by placing the tubing underneath one side and<br />

ensuring an adequate seal. Use a two-hand technique to maximise<br />

the seal with the patient’s face (Fig. 4.6).<br />

High airway pressures can be generated if the tidal volume or<br />

inspiratory flow is excessive, predisposing to gastric inflation and<br />

subsequent risk of regurgitation and pulmonary aspiration. The<br />

possibility of gastric inflation is increased by:<br />

• malalignment of the head and neck, and an obstructed airway;<br />

• an incompetent oesophageal sphincter (present in all patients<br />

with cardiac arrest);<br />

• a high airway inflation pressure.<br />

Conversely, if inspiratory flow is too low, inspiratory time will<br />

be prolonged and the time available to give chest compressions<br />

is reduced. Deliver each breath over approximately 1 s and transfer<br />

a volume that corresponds to normal chest movement; this<br />

represents a compromise between giving an adequate volume,<br />

minimising the risk of gastric inflation, and allowing adequate time<br />

<strong>for</strong> chest compressions. During CPR with an unprotected airway,<br />

Fig. 4.6. Mouth-to-mask ventilation.<br />

give two ventilations after each sequence of 30 chest compressions.<br />

Self-inflating bag<br />

The self-inflating bag can be connected to a facemask, tracheal<br />

tube or supraglottic airway device (SAD). Without supplemental<br />

oxygen, the self-inflating bag ventilates the patient’s lungs with<br />

ambient air (21% oxygen). The delivered oxygen concentration can<br />

be increased to about 85% by using a reservoir system and attaching<br />

oxygen at a flow 10 l min −1 .<br />

Although the bag-mask device enables ventilation with high<br />

concentrations of oxygen, its use by a single person requires considerable<br />

skill. When used with a face mask, it is often difficult to<br />

achieve a gas-tight seal between the mask and the patient’s face,<br />

and to maintain a patent airway with one hand while squeezing<br />

the bag with the other. 330 Any significant leak will cause hypoventilation<br />

and, if the airway is not patent, gas may be <strong>for</strong>ced into<br />

the stomach. 331,332 This will reduce ventilation further and greatly<br />

increase the risk of regurgitation and aspiration. 333 Cricoid pressure<br />

can reduce this risk 334,335 but requires the presence of a trained<br />

assistant. Poorly applied cricoid pressure may make it more difficult<br />

to ventilate the patient’s lungs. 334,336–339<br />

The two-person technique <strong>for</strong> bag-mask ventilation is preferable<br />

(Fig. 4.7). One person holds the facemask in place using a jaw<br />

thrust with both hands, and an assistant squeezes the bag. In this<br />

way, a better seal can be achieved and the patient’s lungs can be<br />

ventilated more effectively and safely.<br />

Once a tracheal tube or a supraglottic airway device has been<br />

inserted, ventilate the lungs at a rate of 10 breaths min −1 and continue<br />

chest compressions without pausing during ventilations. The<br />

laryngeal seal achieved with a supraglottic airway device is unlikely<br />

to be good enough to prevent at least some gas leaking when inspiration<br />

coincides with chest compressions. Moderate gas leakage is<br />

acceptable, particularly as most of this gas will pass up through the<br />

patient’s mouth. If excessive gas leakage results in inadequate ventilation<br />

of the patient’s lungs, chest compressions will have to be

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!