Protecting the measuring site<br />

6 Protecting the measuring site<br />

6.1 Theft and vandalism<br />

The site should be protected against theft and vandalism as well as<br />

against any farming or field work. Therefore, the site should be<br />

fenced and signposts could give information about the purpose of the<br />

site.<br />

6.2 Cable protection<br />

Cables should be protected against rodents with plastic protection<br />

tubes. <strong>UMS</strong> offers dividable protection tubes as accessory. For long<br />

term studies we recommend to dig cables a few centimeters below<br />

soil surface inside protection tubes<br />

6.3 Frost<br />

6.3.1 Protection against frost<br />

Tensiometer are filled with water and therefore are endangered by<br />

frost!<br />

Do not store filled Tensiometer at temperatures below–5°C. Do<br />

not leave filled Tensiometers over night in your car, in a measuring<br />

hut, etc.<br />

Do not fill the Tensiometers with Ethanol, as this is corrosive for<br />

some materials (i. e. PMMA) and will destroy these.<br />

Also it is not recommended to fill the Tensiometers with Decalin,<br />

mono-ethylene-glycol, di-ethylene-glycol, etc. These could harm any<br />

of the materials, destroy the ceramic cup or leak into the soil.<br />

<strong>T8</strong> Tensiometer may remain installed during the winter if the cup is<br />

positioned in a depth of at least 20 cm. Then, the frost will ingress<br />

the cup slowly without damaging the pressure sensor. The reading<br />

will jump to a constant value. After unfreezing the Tensiometer will<br />

continue to work.<br />