Protecting the measuring site<br />
6 Protecting the measuring site<br />
6.1 Theft and vandalism<br />
The site should be protected against theft and vandalism as well as<br />
against any farming or field work. Therefore, the site should be<br />
fenced and signposts could give information about the purpose of the<br />
site.<br />
6.2 Cable protection<br />
Cables should be protected against rodents with plastic protection<br />
tubes. <strong>UMS</strong> offers dividable protection tubes as accessory. For long<br />
term studies we recommend to dig cables a few centimeters below<br />
soil surface inside protection tubes<br />
6.3 Frost<br />
6.3.1 Protection against frost<br />
Tensiometer are filled with water and therefore are endangered by<br />
frost!<br />
Do not store filled Tensiometer at temperatures below–5°C. Do<br />
not leave filled Tensiometers over night in your car, in a measuring<br />
hut, etc.<br />
Do not fill the Tensiometers with Ethanol, as this is corrosive for<br />
some materials (i. e. PMMA) and will destroy these.<br />
Also it is not recommended to fill the Tensiometers with Decalin,<br />
mono-ethylene-glycol, di-ethylene-glycol, etc. These could harm any<br />
of the materials, destroy the ceramic cup or leak into the soil.<br />
<strong>T8</strong> Tensiometer may remain installed during the winter if the cup is<br />
positioned in a depth of at least 20 cm. Then, the frost will ingress<br />
the cup slowly without damaging the pressure sensor. The reading<br />
will jump to a constant value. After unfreezing the Tensiometer will<br />
continue to work.<br />
41/56