3.5 Connecting <strong>T8</strong><br />
3.5.1 Spot reading with the INFIELD7 handheld<br />
The <strong>T8</strong> is fitted with an 8-pin plug. The plug can be connected to an<br />
INFIELD7 handheld measuring device. The INFIELD7 displays and<br />
stores the soil water tension, the soil temperature and the filling<br />
status.<br />
3.5.2 Connection to a data logger<br />
For connecting <strong>T8</strong> to a data logger or any other data acquisition<br />
devices connecting and extension cables are required. Find <strong>T8</strong><br />
cables in the chapter “Accessories”.<br />
The supply voltage does not has to be stabilized.<br />
If the Tensiometer is not permanently powered the warm-up<br />
before a measurement should be at least 10 seconds.<br />
The Tensiometer plug should be covered with the supplied<br />
protective cap anytime the <strong>T8</strong> is not connected to an extension<br />
cable.<br />
Never connect the signal outputs of the <strong>T8</strong> to a supply voltage<br />
source!<br />
The supply voltage must not exceed 18 VDC.<br />
Tightly screw together all plugs of connecting cables (CC-8/...) or<br />
extension cables (EC-8/...). Do this again after a few minutes as<br />
only then the connections will be absolutely water proof.<br />