Ascorbic Acid on<br />

Allure <strong>Acidix</strong> (LC/MS)<br />

Peak List:<br />

1. unknown (2.33 min)<br />

2. ascorbic acid (8.84 min)<br />

Sample:<br />

Inj.: 5µL<br />

Conc.: 50µg/mL<br />

Solvent: water:methanol (1:1, v/v)<br />

2<br />

Application Note: LC_0114<br />

<strong>Column</strong>: Allure <strong>Acidix</strong><br />

Catalog #: 9162565<br />

Dimensions: 150 x 4.6mm<br />

Particle Size: 5µm<br />

Pore Size: 60Å<br />

Conditions:<br />

Mobile Phase:<br />

Flow:<br />

Temp.:<br />

Det.:<br />

Interface:<br />

Ion Mode:<br />

5 mM ammonium formate,<br />

pH 3 :methanol (50:50, v/v)<br />

1.0mL/min<br />

ambient<br />

PE/Sciex API 150 EX<br />

Heated Nebulizer<br />

negative<br />

Needle current: 4µA<br />

Temp.: 350°C<br />

Orifice: -30V<br />

Ring: -300V<br />

Auxiliary gas: 6000 cc/min<br />

1<br />

0 1 2 3 4 5 6 7 8 9 min.<br />

Restek Corporation 110 Benner Circle Bellefonte, PA 16823 814-353-1300 800-356-1688 Fax: 814-353-1309