<strong>Restek</strong> <strong>Product</strong> <strong>Instruction</strong> Sheet -- ECD/FID Make-up Gas Kit<br />
4. Remove any existing make-up gas fittings or any inserts inside the 1/4" detector port.<br />
5. Route the 1/16" line through an existing oven hole to the variable flow outlet. If no hole exists,<br />
one will have to be drilled. Be sure to drill the hole in a place away from existing lines or<br />
electronics. Before pushing the tubing through the insulation, place a plastic cap or a piece of tape<br />
over the tubing end.<br />
6. Connect the 1/16" line to the variable flow restrictor gas outlet fitting.<br />
7. Insert the Silcosteel® guide (item 6). (Omit the Silcosteel guide if an FID will be used.) Connect<br />
the 1/4" make-up gas fitting (item 1) to the detector port using the 1/4" nut and ferrule (items 2 &<br />
3) provided.<br />
8. Cap off the column connection to the detector and leak-check all connections with <strong>Restek</strong>'s Leak<br />
Detective (cat.# 21607) or an equivalent leak detector. After confirming that all fittings are leakfree,<br />
adjust the variable flow restrictor to set the make-up gas flow as indicated in the instrument<br />
manual.<br />
9. Install the column using the 1/16" nut and ferrule (items 4 & 5) according to the procedure<br />
outlined in your instrument manual and <strong>Restek</strong>'s Capillary Column Installation Guide.<br />
* We highly re<strong>com</strong>mend using our heat treated instrument grade SS tubing (cat.# 21512) for plumbing<br />
the make-up gas system to avoid the possibility of contamination.<br />
Contact <strong>Restek</strong> Technical Service (or an <strong>au</strong>thorized distributor) if you have questions about the ECD/FID<br />
Make-up Gas Kit or any other <strong>Restek</strong> product.<br />
Australian Distributor - Chromalytic Technology Pty Ltd - Tel (03) 9762 2034 Fax (03) 9761169 www<br />
Copyright © 1998 <strong>Restek</strong> Corporation. All Rights Reserved.<br />