COBAS ACCURACY<br />

Apparatus: Roche Cobas e601<br />

Analyte: <strong>PROCALCITONIN</strong><br />

Samples: PC PCT Level 1 & 2 (Roche QC)/VIDAS QC 1&2 Unit: ug/L<br />

Operator: Jennifer Annand Accuracy tolerance limit : 10.0%<br />

Level Sample Id Date Result 1 Result 2 Mean Result Assigned Deviation Accepted<br />

Mean (%)<br />

1 Vidas QC1 15-Jun-09 12.79 13.05 12.92 16.80 -23.1% NO<br />

2 Vidas QC2 15-Jun-09 1.40 1.43 1.42 1.79 -20.9% NO<br />

3<br />

4 PC PCT 1 15-Jun-09 0.494 0.486 0.49 0.48 2.1% YES<br />

5 PC PCT 2 15-Jun-09 9.80 9.88 9.84 9.69 1.5% YES<br />

Level Sample Id Date Result 1 Result 2 Mean Result Assigned<br />

Mean<br />

Deviation<br />

(%)<br />

Accepted<br />

1 Vidas QC1 16-Jun-09 12.85 12.88 12.87 16.80 -23.4% NO<br />

2 Vidas QC2 16-Jun-09 1.45 1.40 1.43 1.79 -20.4% NO<br />

3 FALSE<br />

4 PC PCT 1 16-Jun-09 0.477 0.485 0.48 0.48 0.2% YES<br />

5 PC PCT 2 16-Jun-09 9.71 9.82 9.77 9.69 0.8% YES