The talar window trial is removed, and the keel broach is used to<br />
finish preparation of the keel slot. (FIGURE 38) When the keel broach<br />
is properly inserted to depth its angled top surface will align in the<br />
same plane as the posterior cut and the vertical face lines up with<br />
the edge formed by the intersection of the anterior and superior cut<br />
surfaces. (FIGURE 39)<br />
Replace the talar window trial to protect the talar cut surface during<br />
the final tibial preparation.<br />
Figure 38<br />
Figure 39<br />
Figure 40<br />