25.09.2014 Views

PGT INSTRUmENTATIoN - Small Bone Innovations

PGT INSTRUmENTATIoN - Small Bone Innovations

PGT INSTRUmENTATIoN - Small Bone Innovations

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

The talar window trial is used to check that all talar cuts are complete<br />

and accurate. (FIGURES 34a+b) If the window trial does not fit, the<br />

bone rasp is used to remove excess unresected bone. (FIGURE 37)<br />

Preparation of the Keel<br />

Place two 2.4 mm pins in the anterior holes on talar window trial<br />

for additional stability during keel preparation. (FIGURE 35) The<br />

straight keel mill (FIGURE 36) is placed in the center slot of the<br />

window trial. The mill can be utilized with a drill or wire driver. The<br />

straight mill is pushed into the talar surface until it is flush with the<br />

stop on the mill and moved anteriorly on the slot. (FIGURES 37a+b)<br />

Slot for mill and broach<br />

Rasp<br />

Figure 33<br />

Pin holes<br />

Window Trial<br />

Figure 34a<br />

Figure 34b<br />

Keel Mill<br />

Figure 36<br />

Figure 35<br />

Figure 37a<br />

Figure 37b<br />

14<br />

S.T.A.R. ® Scandinavian Total Ankle Replacement Surgical Technique

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!