BOD analyzes were carried out using a BOD incubator (Hach, model 205) which is especially<br />

designed for BOD tests. The sampling bottles are connected with a tube (for oxygen<br />

measurements) <strong>and</strong> put in a compartment that keeps the temperature at 20 °C. A computer<br />

records the oxygen consumption over time. The procedure was performed according to the<br />

product manual. One benefit of the BOD incubator is that it allows reading of the oxygen<br />

consumption (BOD) continuously (instead of only a final measurement after 5 days as the<br />

“manual” procedure described above). For calculations of the k-value (Appendix E) readings<br />

were recorded with 0.5 days intervals (Appendix F).<br />

One full testing event with the BOD incubator was successful. Other testing events failed due to<br />

technical problems with the BOD incubator or lack of available storage space in the fridge where<br />

the incubator was.<br />

Manual method<br />

The “manual” measuring of BOD 5 is done with the following procedure:<br />

‐ Filing 300 ml airtight bottles with diluted waste water (The purpose of diluting the<br />

samples was to reach a DO above 7 mg O 2 / liter, which may be necessary for a<br />

successful BOD reading after 5 days, as the DO must be above 2 mg O 2 / liter at final<br />

reading. Dilution 1:1 with distilled water was enough for this purpose).<br />

‐ Measuring the initial DO level in the samples with a DO meter.<br />

‐ Storing the bottles in a refrigerator at 20 °C for 5 days.<br />

‐ Measuring the final DO with a DO meter.<br />


<br />

The BOD 5 was calculated as the decrease of DO per liter of water.<br />

This method was considered as slightly unreliable as the reading of the DO meters was difficult.<br />

The DO measured DO concentration was not stable over time, hence making it difficult to<br />

measure the final DO.<br />