Part # <strong>35345</strong> Revised 11/02


PATIENT REPORT IMAGES STORAGE MANAGEMENT<br />

for<br />

Digital Colposcopy Systems<br />

U S E R ’ S<br />

G U I D E


PATIENT REPORT IMAGES STORAGE MANAGEMENT<br />

for<br />

Digital Colposcopy Systems<br />

U S E R ’ S<br />

G U I D E<br />

LEISEGANG 2000 ©<br />

Part # <strong>35345</strong> Revised 11/02


Contents<br />

Table of Contents<br />

Leisegang User Assistance Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II<br />

Preface<br />

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III<br />

Summary of Warnings & Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV<br />

Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VI<br />

General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . VIII<br />

Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IX<br />

Symbols & Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . X<br />

Section 1 ■ How to Use the Prism Digital Colposcopy System<br />

A Quick Overview of a Typical Colposcopy Visit . . . . . . . . . . . . 1.1<br />

Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2<br />

Section II ■ Clinicians<br />

Attending Clinician Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1<br />

Section III ■ Patients<br />

Selecting Patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1<br />

Patient Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2<br />

Section IV ■ Imaging<br />

Understanding the Imaging Screen . . . . . . . . . . . . . . . . . . . . . . 4.1<br />

Patient Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.9<br />

Section V<br />

■ System Backup & Restoration<br />

Using Automatic Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1<br />

Using Manual Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.3<br />

Section VI ■ System Maintenance<br />

Accessing System Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

I


Preface<br />

USER ASSISTANCE<br />

Leisegang User Assistance Information<br />

<strong>CooperSurgical</strong>/Leisegang Technical Support<br />

Phone: (800) 243-2974<br />

Fax: (800) 262-0105<br />

E-mail: techsupport@coopersurgical.com<br />

Please have the following information<br />

available before you call:<br />

Software Version Number: __________________<br />

The software version number can be found on the<br />

Introduction screen.<br />

II<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


INTRODUCTION<br />

Preface<br />

Preface<br />

Thank you for purchasing the Leisegang Prism <br />

Digital Colposcopy System<br />

Purpose of the Leisegang Prism <br />

Digital Colposcopy System<br />

The Leisegang Prism Digital Colposcopy System is indicated for use on<br />

patients who received an abnormal cervical cytology result or who are<br />

suspected to have a lower genital tract infection. The Leisegang Prism Digital<br />

Colposcopy System will provide magnified digital image documentation of<br />

vaginal, cervical, and external genital tissue examinations. The system may<br />

also be used as a visual aid for related gynecological procedures. The system<br />

is intended for visual examination and documentation purposes only.<br />

Purpose of the Manual<br />

The manual will provide you with operating instructions and information on<br />

the Leisegang Prism Digital Colposcopy System with Image Viewer software.<br />

Please read through the manual carefully. All persons using the Leisegang<br />

Prism Digital Colposcopy System should be trained in colposcopy. The<br />

contents of the manual are not intended to be a replacement or<br />

substitute for certified medical training.<br />

Note: The material outlined in this manual should be reviewed and understood<br />

prior to operation of the Leisegang Prism Digital Colposcopy System.<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

III


Preface<br />

WARNINGS & CAUTIONS<br />

Summary of Warnings and Cautions<br />

A Warning or Caution normally includes the following:<br />

■<br />

■<br />

■<br />

■<br />

The signal word, “WARNING” or “CAUTION”.<br />

A hazard statement, which briefly explains the problem.<br />

The consequences if instructions are not followed, to give<br />

the user a clear idea of the risk.<br />

Instructions explaining the do’s and don’ts necessary<br />

to avoid the hazard.<br />

See the Sample Warning, Caution, and Note Below<br />

■<br />

■<br />

■<br />

WARNING A Warning is a statement that alerts the user to the<br />

possibility of injury, death, or other serious adverse reaction<br />

associated with the use or misuse of the device.<br />

CAUTION A Caution is a statement that alerts the user to the<br />

possibility of a problem with the device associated with its use or<br />

misuse. Such problems include device malfunction, device failure,<br />

damage to the device, or damage to other property. The caution<br />

statement includes the precaution that should be taken to avoid<br />

the hazard.<br />

NOTE Important notes are emphasized this way to give<br />

supplemental information.<br />

The following is a summary of all Warnings and Cautions associated with<br />

the Leisegang Prism Digital Colposcopy System. These Warnings and<br />

Cautions also appear elsewhere in this User’s Guide and are<br />

summarized here for reader convenience.<br />

■<br />

■<br />

■<br />

WARNING Mains isolation can only be achieved by disconnecting<br />

power supply cord from wall receptacle.<br />

WARNING DO NOT use a converter adapter that will convert the<br />

three-prong AC plug to a two-prong line plug. The power supply in<br />

the Leisegang Prism Digital Colposcopy System will not be properly<br />

grounded and electric shock may result. Use only a Leisegang<br />

supplied power cord.<br />

WARNING NEVER operate the Leisegang Prism Digital Colposcopy<br />

System in the presence of flammable vapors or explosive gases,<br />

chemicals, or in areas where these materials are commonly used.<br />

IV<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


WARNINGS & CAUTIONS ■ Continued<br />

Preface<br />

Warnings and Cautions continued<br />

■<br />

■<br />

■<br />

■<br />

■<br />

■<br />

■<br />

■<br />

WARNING DO NOT disassemble. Disassembly of the Leisegang Prism <br />

Digital Colposcopy System could expose electrical hazards.<br />

Only Leisegang qualified and trained maintenance personnel are<br />

authorized to service the colposcope. Disassembly or modification of the<br />

hardware is a violation of the warranty.<br />

WARNING Grounding reliability can only be achieved when equipment is<br />

connected to a receptacle marked “Hospital Only” or “Hospital Grade”.<br />

WARNING This equipment shall not be stored in an environment<br />

exceeding the following:<br />

— Ambient temperature range of -40°C to 70°C<br />

— Relative humidity range of 10% to 100%<br />

— Atmospheric pressure range of 50 kilopascals (KPA) to 106 KPA<br />

CAUTION DO NOT touch the optical or illumination lenses except<br />

as described in the maintenance section of this manual.<br />

CAUTION DO NOT spray cleaning solution directly onto<br />

any part of the unit.<br />

CAUTION Keep all liquids away from electrical equipment<br />

to avoid the possibility of shock and device damage.<br />

CAUTION Occasionally inspect the power cord for signs of cuts or dents.<br />

CAUTION Avoid extreme environmental conditions when storing<br />

or operating the Leisegang Prism Digital Colposcopy System.<br />

Avoid areas where it could get wet or be exposed to extreme<br />

temperature or humidity, direct sunlight and dust.<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

V


Preface<br />

REGULATORY INFORMATION<br />

Regulatory Information<br />

Model Name Leisegang Prism Digital Colposcopy System<br />

Model Number Prism <br />

Manufacturer <strong>CooperSurgical</strong>, Inc., Trumbull, CT 06611<br />

203-601-5200 Fax: 800-262-0105<br />

Product Classification<br />

IPXO (Ordinary equipment with no special protection against ingress of water.)<br />

Electrical Rating 120 VAC, 2A<br />

Degree of Protection against Electrical Shock Type B equipment.<br />

Fuses UL Listed 2A, 250Vac, Sloblo (Qty. 2)<br />

Protective Packing and Storage Constraints<br />

CAUTION The Leisegang Prism Digital Colposcopy System contains<br />

sensitive electronic equipment. Handle with care. Any external damage to<br />

container occurring during transportation should be reported immediately<br />

upon receipt.<br />

Technical Description<br />

The Leisegang Prism Digital Colposcopy System is comprised of<br />

a computer with the Prism System software, a VGA display monitor, printer,<br />

keyboard, mouse, and digital camera which when used in conjunction with a<br />

Leisegang standard optical colposcope serves as a documentation-enhancing<br />

device. Outlets for each of the AC powered devices are located behind the<br />

access panel on the lower rear of the cart. The supplied hospital-grade power<br />

cord connects to the AC power inlet and must be plugged into a hospital-grade<br />

wall outlet.<br />

The Leisegang Prism Digital Colposcopy System has been tested<br />

and found to comply with the limits established for digital medical<br />

equipment regarding electromagnetic compatibility and avoidance of<br />

electromagnetic interference, pursuant to IEC 601-1. If not installed and<br />

operated in accordance with the manufacturer’s instructions, this equipment<br />

may cause harmful interference to radio communications and other medical<br />

equipment. Change or modification to this equipment without the expressed<br />

written approval of Leisegang is strictly prohibited and could void<br />

the user’s authority to operate this equipment.<br />

VI<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


REGULATORY INFORMATION ■ Continued<br />

Preface<br />

Regulatory Information continued<br />

1 The Leisegang Prism Digital Colposcopy System is only authorized for use<br />

with the provided computer assembly, which has all of the neccessary hardware<br />

and software items factory installed and configured. The computer<br />

must be plugged into the system AC outlets and be properly secured to<br />

the lower deck using the included hardware. No other computer installation<br />

is authorized.<br />

2 The Leisegang Prism Digital Colposcopy System is only authorized for<br />

connection of the provided computer display, which must be properly secured<br />

to the upper deck using the included hardware. The display power cord must<br />

be plugged in to the system AC outlets. No other display installation is<br />

authorized.<br />

3 The Leisegang Prism Digital Colposcopy System is only authorized for use<br />

with UL1950 compliant printers. The printer must be physically located on<br />

the lower deck. The printer power cord must be plugged into the system AC<br />

outlets. No other printer installation is authorized.<br />

4 The use of accessory equipment that does not comply with the equivalent<br />

safety requirements of this equipment may lead to a reduced level of safety<br />

of the resulting system. Consideration of the choice shall include:<br />

■<br />

■<br />

Use of the accessory in the patient vicinity<br />

Evidence that the safety certification of the accessory has been<br />

performed in accordance to the appropriate IEC 601 and/or<br />

IEC 601-1-1 harmonized national standard.<br />

Contact the Leisegang Customer Service Department at (800) 243-2974<br />

for a current list of authorized printers.<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

VII


Preface<br />

GENERAL INFORMATION<br />

General Product Information<br />

Technical Support<br />

Technical Support for the Leisegang Prism Digital Colposcopy System is<br />

available by contacting Leisegang at (800) 243-2974. Repairs, upgrades, and/or<br />

modifications to this equipment may only be made by Leisegang<br />

authorized personnel.<br />

User Detachable Parts<br />

AC Power Cord Hospital grade AC power cords come standard with the<br />

Leisegang Prism Digital Colposcopy System. The Leisegang Prism Digital<br />

Colposcopy System should only be used with the Leisegang supplied hospitalgrade<br />

AC power cord. The cord attaches to the AC power inlet located on the<br />

lower rear of the cart. The other end of the power cord attaches to a hospitalgrade<br />

wall outlet supplying the appropriate power for the entire system.<br />

Printer One printer comes standard with the Leisegang Prism Digital<br />

Colposcopy System and is connected at the rear of the computer using<br />

the supplied printer data cable. The ink cartridges may require peridic<br />

replacement. Refer to the printer User Guide for information on replacing<br />

ink cartridges.<br />

Keyboard One keyboard comes standard with the Leisegang Prism Digital<br />

Colposcopy System and is connected at the rear of the computer.<br />

Mouse One mouse comes standard with the Leisegang Prism Digital<br />

Colposcopy System and is connected at the rear of the computer.<br />

No other user detachable parts are included in the Leisegang Prism Digital<br />

Colposcopy System installation. Any other attaching or detaching of parts to<br />

the Leisegang Prism Digital Colposcopy System should only be attempted by<br />

Leisegang authorized service personnel.<br />

User Serviceable Parts<br />

There are no user serviceable parts in the Leisegang Prism Digital Colposcopy<br />

System. Repairs should only be performed by authorized service personnel.<br />

Routine Maintenance<br />

Routine Maintenance for the Leisegang Prism Digital Colposcopy System<br />

consists of periodic cleaning that is accomplished (excluding lenses) by wiping<br />

down the equipment with a cloth, slightly dampened with a mild disinfectant.<br />

The lenses should be cleaned with either 70% isopropyl alcohol, or any<br />

commercially available lens or glass cleaner. The period of cleaning is<br />

at the user’s discretion.<br />

VIII<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


PARTS<br />

Preface<br />

Parts<br />

Floppy Disk<br />

On/Off<br />

Switch<br />

USB Ports<br />

Computer ■ Front View<br />

Monitor<br />

Camera<br />

Network<br />

Keyboard<br />

Mouse<br />

Serial Port<br />

Power<br />

Printer<br />

Computer ■ Rear View<br />

Display<br />

Screen<br />

Power<br />

Button<br />

CRT Monitor ■ Front View<br />

User’s Guide<br />

<strong>CooperSurgical</strong> CS1600 Digital Colposcopy System<br />

IX


Preface<br />

CAUTIONS<br />

Product Labeling, Warning and<br />

Cautionary Symbols and Definitions<br />

S Y M B O L<br />

D E F I N I T I O N<br />

When found anywhere on the outside of the<br />

Leisegang Prism Digital Colposcopy System<br />

this symbol means “Caution–Refer to User’s<br />

Guide for more information.”<br />

Warning: No userserviceable<br />

parts inside.<br />

Repairs and upgrades<br />

should only be performed<br />

by <strong>CooperSurgical</strong><br />

authorized personnel.<br />

No one except <strong>CooperSurgical</strong> authorized<br />

technical staff should attempt to open any<br />

part of the Leisegang Prism Digital<br />

Colposcopy System. Failure to comply<br />

could be dangerous and may result in<br />

loss of warranty.<br />

Located on the inside of the equipment,<br />

accessible only to authorized service<br />

personnel, this symbol identifies an electrical<br />

connection to earth ground through the AC<br />

power plug.<br />

Type “B” equipment<br />

X<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


THE BASICS<br />

Section I<br />

How to Use the Prism Digital Colposcopy System<br />

A Quick Overview of a Typical Colposcopy Visit<br />

1 Turn on the power on the Prism Digital Colposcopy System,<br />

computer, display and light source.<br />

2 On the Welcome Screen, select the Continue button.<br />

3 Select a Physician.<br />

4 Enter Patient Name, ID, and Birth Date. Select Start Visit button.<br />

5 Focus the scope and select the Imaging button. On the Imaging<br />

screen, select the Capture button to take pictures.<br />

6 If desired, review the images from the current visit, or compare<br />

them to images from previous visits.<br />

7 Select End Visit and indicate whether the Colposcopy Report<br />

should be printed.<br />

8 Continue with a new Patient Visit, or<br />

9 Log off the Graphic User Interface software and turn off power<br />

to the Prism Digital Colposcopy System.<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

1.1


Section I<br />

THE BASICS ■ Continued<br />

Getting Started<br />

■<br />

Turn the system on by pushing the On/Off buttons located<br />

on the computer.<br />

NOTE:<br />

A series of computer diagnostic messages will appear on the screen during<br />

the system boot-up before the Welcome screen appears. Disregard these<br />

messages, and wait for the Welcome screen.<br />

Welcome Screen/PRISM <br />

■<br />

■<br />

Click on the Continue button on the bottom of the Welcome<br />

screen or anywhere in the Welcome screen except the Power<br />

Down button.<br />

Click on the Exit to Windows button to utilize the<br />

Windows Operating System.<br />

1.2<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


THE BASICS<br />

Section I<br />

How to Use the Prism Digital Colposcopy System continued<br />

After you click on the Continue button, a Clinician List screen will appear.<br />

■<br />

Selecting the Logout button on the Clinician List screen will return you<br />

to the Welcome screen.<br />

Attending Clinician Screen<br />

NOTE:<br />

All names used to describe the Prism Software Program , including physicians<br />

and patients, are fictitious. Any similarities found in this manual are coincidental.<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

1.3


Section I<br />

THE BASICS ■ Continued<br />

How to Use the Prism Digital Colposcopy System continued<br />

The Leisegand Prism Digital Colposcopy System comes with a Security<br />

Password feature, which provides a level of protection for your patient’s<br />

records. The Security Password feature can be “enabled” or “disabled”<br />

at any time.<br />

To enable the Security Password feature, do the following:<br />

■<br />

■<br />

■<br />

Enter the System Maintenance screen.<br />

Click on the Computer System tab, then the Security button.<br />

Under Authorization Code, enter the code provided to you in the<br />

Prism inclusion package. The code will change each month.<br />

NOTE:<br />

■<br />

■<br />

The Current Authorization Code can be provided to you when you call<br />

<strong>CooperSurgical</strong>/Leisegang Customer Service or Technical Support.<br />

Check Password Enabled or leave unchecked if password security<br />

is not desired.<br />

Click OK first, then click Attending Clinician button to return<br />

to Attending Clinician List screen.<br />

NOTE:<br />

NOTE:<br />

As shipped from the factory, the system is set to Password Disabled.<br />

Security Enabled disallows most System Maintenance functions<br />

except security.<br />

Enable Security ■ System Maintenance Screen<br />

1.4<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


CLINICIANS<br />

Section II<br />

Clinicians<br />

Attending Clinician Screen<br />

Adding a Clinician’s Name to the List<br />

To add a clinician’s name to the Attending Clinician list, do the following—<br />

Attending Clinician Screen<br />

Security Password Disabled<br />

■<br />

■<br />

Click on the Add button.<br />

The Add Physician window will appear.<br />

Add Physician Screen<br />

Security Password Enabled<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

2.1


Section II<br />

CLINICIANS ■ Continued<br />

Attending Clinician Screen continued<br />

■<br />

■<br />

■<br />

Input an “ID” which can be a nickname, first name or perhaps a title.<br />

The “ID” field will appear on the Name button in the software<br />

program when that particular MD is logged in. The ID field allows up to<br />

12 characters, which is the field limitation on the software button.<br />

Enter the first, middle and last name of the clinician into the “Name”<br />

field. The name entered here will appear on all colposcopy reports.<br />

Enter a “Password”. The “Password” must be entered for the OK button<br />

to work. Once entered, the “Password” will remain inactive unless the<br />

password security feature is “enabled” in the System Maintenance screen.<br />

NOTE: You may want to<br />

write your password<br />

down and store it<br />

in a safe place for<br />

future reference.<br />

Attending Clinician Screen<br />

Security Password Enabled<br />

■<br />

■<br />

■<br />

■<br />

If the Password Security Feature is “enabled”, then each time a<br />

clinician enters the system, they will have to enter the appropriate<br />

password associated with their name to enter patient files.<br />

When the Password Security Feature is “enabled”, Add, Delete and Edit<br />

functions are not allowed until a valid password is entered for the<br />

selected clinician.<br />

You can tell when the Password Security feature is “enabled” when<br />

the Password prompt is present.<br />

When the Password Security feature is “disabled”, “Password not used”<br />

appears at the top of the screen.<br />

2.2<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


CLINICIANS ■ Continued<br />

Section II<br />

Attending Clinician Screen continued<br />

Attending Clinician’s Name Already on List<br />

■<br />

■<br />

■<br />

■<br />

If the appropriate physician’s name is displayed in the Attending<br />

Clinician List, manipulate the Navigator button until the cursor is<br />

over the name.<br />

Click on the Mouse Select button and the name will be highlighted<br />

in dark blue.<br />

The password will always be “OK” when the “password” is enabled<br />

until changed by the selected physician.<br />

Then proceed to the “Selecting Patients” section in this manual.<br />

Attending Clinician Screen<br />

Security Password Disabled<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

2.3


Section II<br />

CLINICIANS ■ Continued<br />

Attending Clinician Screen continued<br />

Deleting a Clinician’s Name From the Clinician List with<br />

the Security Password Feature Disabled<br />

■<br />

■<br />

Click on the Clinician’s name you would like to delete.<br />

The clinician’s name will be highlighted in blue after you click on it.<br />

Attending Clinician Screen<br />

Security Password Disabled<br />

■<br />

Click on the Delete Clinician button at the bottom middle of the screen.<br />

NOTE: To delete a clinician name while the Security Password feature is “enabled”<br />

you must first enter the correct password.<br />

2.4<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


CLINICIANS ■ Continued<br />

Section II<br />

Attending Clinician Screen continued<br />

■<br />

■<br />

A window will appear asking if you want to delete that Attending<br />

Clinician’s name.<br />

Indicating Yes will delete the clinician’s name from the list.<br />

Delete Clinician Screen<br />

Security Password Enabled<br />

■<br />

The Attending Clinician List will reappear with the<br />

clinician’s name deleted.<br />

NOTE: All patient information is retained in the system after physician is deleted.<br />

Attending Clinician Screen<br />

Security Password Enabled<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

2.5


Section II<br />

CLINICIANS ■ Continued<br />

Attending Clinician Screen continued<br />

Editing a Clinician’s Name on the Attending Clinician List<br />

■<br />

■<br />

■<br />

■<br />

With the appropriate password entered (if password security is<br />

enabled), click on the clinician’s name you would like to edit.<br />

The clinicians name will be highlighted in blue after you click on it.<br />

Click on the Edit button on the bottom of the screen.<br />

Make selected edits then click OK to execute or Cancel to refuse<br />

any changes.<br />

Edit Clinician Name Screen<br />

Security Password Enabled<br />

NOTE: To edit a clinician’s name when the password is “enabled” you<br />

must first enter the correct password.<br />

2.6<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


PATIENTS<br />

Section III<br />

Patients<br />

Selecting Patients<br />

1 Click on the name of the clinician that the patient is seeing<br />

from the Attending Clinician List.<br />

2 If you do not see the clinician’s name you need and the list is extensive,<br />

then scroll down the list using the scrollbar located to the right of the list.<br />

The clinician’s name will be highlighted in blue after you click on it.<br />

3<br />

1<br />

2<br />

Attending Clinician Screen<br />

Security Password Disabled<br />

3 Click on the Patient button at the top middle of the screen.<br />

The Patient screen will appear. The edit fields will be blank.<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

3.1


Section III<br />

PATIENTS ■ Continued<br />

Patients continued<br />

Patient Screen<br />

New Patient<br />

If your patient has not had a previous examination with your Prism Digital<br />

Colposcopy System, add her to the system in the following way:<br />

1 Type the patient’s name in the three edit fields (Last Name, First Name,<br />

Middle Initial) using the regular keyboard.<br />

2 Type the ID number in the edit field.<br />

3 Type the date of birth in the edit fields, or you can use the Calendar tool.<br />

1<br />

2 3<br />

Patient Screen<br />

NOTE: Adding patient ID is optional but is designed to help you distinguish<br />

between women who may have the same name.<br />

3.2<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


PATIENTS ■ Continued<br />

Section III<br />

Patients continued<br />

Using the Calendar<br />

1 Click on the Calendar icon next to the Birth Date field.<br />

2 Click on the selected date on the calendar.<br />

3 Use the arrows on the calendar to scroll to the correct month and year.<br />

1<br />

Patient Screen<br />

3<br />

2<br />

NOTE:<br />

The patient age<br />

is calculated after the<br />

birth date is entered.<br />

The calculated age is<br />

also updated automatically<br />

for all new visits.<br />

Calendar Screen<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

3.3


Section III<br />

PATIENTS ■ Continued<br />

Patients continued<br />

Existing Patient<br />

If your patient has had a previous examination with your Prism Digital<br />

Colposcopy System, you can bring her records onscreen in one of two ways:<br />

1 Click on the Patient List button in the top<br />

middle of the Patient screen.<br />

A Patient List screen will appear with a list of<br />

patients who have already been entered into the system.<br />

■<br />

■<br />

■<br />

Scroll through the list using the scrollbar on the right if the list is<br />

extensive.<br />

Click on the appropriate patient’s name, then click OK to Select.<br />

Alternatively, you could “double click” on the patient’s name to display.<br />

The patient’s name will be highlighted in yellow.<br />

1<br />

NOTE: Double clicking on the<br />

column descriptor will<br />

sort the Patient List<br />

screen.<br />

Patient List Screen<br />

2 Start typing the patient’s name in the edit field. If the Auto search feature is<br />

checked, then the software will perform a search with every letter that is<br />

typed into the patient Last Name box. Typing “S” in the edit field will bring up<br />

the first patient entry whose last name begins with “S.”<br />

3.4<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


PATIENTS ■ Continued<br />

Section III<br />

Patients continued<br />

Type “M” in the edit field to progress to the first patient entry whose name<br />

begins with “SM”. Continue that way until the name of the patient you’re<br />

looking for appears on the screen.<br />

2<br />

NOTE: The Auto Search feature<br />

attempts to guess what<br />

patient in the database is<br />

being selected as you<br />

key in letters of the<br />

patient name.<br />

To disengage the<br />

Auto Search feature,<br />

uncheck the box.<br />

Patient Screen<br />

3<br />

Patient Screen<br />

Start Visit Button<br />

3 Click on the Start Visit<br />

button at the top right of the Patient screen.<br />

After you click on the Start Visit button, the Imaging screen will appear.<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

3.5


Section III<br />

PATIENTS ■ Continued<br />

Notes<br />

3.6<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


IMAGING<br />

Section IV<br />

Imaging<br />

Understanding the Imaging Screen<br />

When you begin an examination, the Today’s Visit filmstrip, representing<br />

the current day’s images, and the Selected Imaging Area will appear empty.<br />

The Previous visit filmstrip may contain visit folders from previous visits.<br />

4<br />

1<br />

3<br />

2<br />

Imaging Screen<br />

The Capture Button<br />

1 The Capture button will initiate the camera to capture an image and place it in<br />

thumbnail size in both the Selected Image area and the Today area.<br />

Imaging Screen Parts<br />

2 The blue bar under the Selected Image area gives the time and image attributes<br />

on the left. The magnification being used is shown on the right. Clicking on<br />

the blue bar will allow the user to add text comments to the image. See Image<br />

Text Entry on page 4.7 for details on how to change the image caption.<br />

3 The blue border in the Today’s Visit filmstrip or the Previous Visit filmstrip<br />

indicates which image is currently in the Image Selected area.<br />

4 The magnification setting should match the step magnification on the<br />

colposcope for accurate documentation.<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

4.1


Section IV<br />

IMAGING ■ Continued<br />

Imaging Screen continued<br />

Delete/Purge Button<br />

■<br />

An image from the<br />

Today’s Visit Filmstrip can be deleted by highlighting it<br />

(so that the blue border is around it) and clicking on the Delete button.<br />

A window will appear asking you if you want to delete that image.<br />

■<br />

Images from Previous visits cannot be deleted, but may be temporarily<br />

purged for the duration of the current visit by clicking on the Purge button<br />

Filter Palette—Digital Green Filter<br />

The Green filter provides for a red-free<br />

green tinted image to better<br />

visualize vascular patterns.<br />

1 Click on the Green filter<br />

button to initiate the<br />

digital green filter.<br />

2 The digital Green<br />

filter allows the user<br />

to view the image<br />

in a green tint, which<br />

highlights vascularity.<br />

3 Click on the Green<br />

filter again to disengage<br />

and return to the<br />

original image.<br />

2<br />

1<br />

Imaging Screen ■ Green Filter<br />

3<br />

4.2<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


IMAGING ■ Continued<br />

Section IV<br />

Imaging Screen continued<br />

Split View Button<br />

Successive clicking on any two thumbnail images will set up the images for<br />

split-view comparison in the Split View screen. The Split View screen is<br />

useful for side-by-side image comparison.<br />

1 Identify the two thumbnail images.<br />

2 Click on the Split View button at the top middle of the screen.<br />

3 The two images you have selected will be the images that appear<br />

in the Split View screen.<br />

4 Click on the Return button to return to the Imaging screen.<br />

2<br />

1<br />

Imaging Screen<br />

3<br />

4<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

Split View Screen<br />

4.3


Section IV<br />

IMAGING ■ Continued<br />

Imaging Screen continued<br />

Contrast Tool<br />

The Contrast tool creates an image with maximum contrast in the red, green,<br />

and blue color planes. The Contrast button can be activated or deactivated<br />

while in the Imaging screen. The Contrast setting can be defaulted to maximum<br />

contrast or no contrast in the System Maintenance Screen User’s Default Tab.<br />

1-2<br />

Imaging Screen ■ Contrast Tool<br />

1 Click on the Contrast button to engage maximum contrast on the image.<br />

2 Click the Contrast button a second time to disengage the contrast<br />

function.<br />

NOTE: The Digital Green Filter is mutually exclusive with Maximum Contrast conditions.<br />

4.4<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


IMAGING ■ Continued<br />

Section IV<br />

Imaging Screen continued<br />

6<br />

4<br />

Focus Button<br />

The Focus button, when activated, engages a unique tool to assist in focusing<br />

and capturing an image. The Focus feature employs a low frame rate video<br />

presentation to provide focusing and color fidelity feedback.<br />

1 Click on the Focus button located at the bottom right of the Imaging screen.<br />

The Focus Mode screen will appear.<br />

NOTE: Today’s and Previous Visit thumbnails are not visible in Focus mode.<br />

Bring the target into focus at higher optical magnification while looking<br />

at the Focus mode screen. When the image is in focus on screen, you should<br />

be in focus through the optics.<br />

2 The camera system<br />

electronics can be<br />

adjusted for Brightness<br />

and Contrast by<br />

engaging the slide<br />

controls at the<br />

bottom of the screen.<br />

3 Click the Freeze<br />

button to save the<br />

camera settings for<br />

Today’s Visit<br />

or<br />

1<br />

4 Click the Default<br />

Imaging Screen<br />

button to return to the<br />

Focus mode settings<br />

and the camera to the<br />

default settings<br />

(established in System<br />

Maintenance page).<br />

5 Click the Cancel<br />

button to negate any<br />

camera settings and<br />

changes and return to<br />

the Imaging screen.<br />

2<br />

3<br />

6 Click the Capture<br />

5<br />

button to capture<br />

Imaging Screen ■ Focus Mode<br />

an image.<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

4.5


Section IV<br />

IMAGING ■ Continued<br />

Imaging Screen continued<br />

Filmstrip Images<br />

1 Patient and Attending Clinician buttons at the top of the Imaging screen<br />

display the names of the physician and the patient, and you may click on these<br />

if you want to return to those screens for information.<br />

2 Previous Visit list shows a folder with the exam date for each of the patient’s<br />

previous visits. The previous list title changes to an exam date after you click<br />

on the appropriate exam folder.<br />

3 Filmstrip—Clicking on a folder will open the file of that visit.<br />

4 Selected Image area—Clicking on any one of those images from a<br />

previous visit will bring its full size version into the Selected Image area.<br />

5 Toggle back to the Previous Visit List (see arrow 2) showing the<br />

Visit Folders by clicking on the filmstrip header.<br />

6 Today filmstrip shows images that have been captured in today’s visit.<br />

The images will be saved and stored in the Visit Folder unless deleted<br />

prior to midnight of the exam date.<br />

7 The Images button on the Imaging Screen prints image reports with blank<br />

comment lines.<br />

1<br />

6<br />

2<br />

4<br />

3<br />

7<br />

Imaging Screen<br />

NOTE:<br />

While engaged in a previous visit image review, any return to the Physician or Patient<br />

sections will reset the images in the Imaging screen back to Today’s images.<br />

4.6<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


IMAGING ■ Continued<br />

Section IV<br />

Imaging Screen continued<br />

Image Text Entry<br />

A text caption appears below the images as they appear in three (3) places:<br />

Image screen and the Image Full screen. Text entry in the caption field can be<br />

changed for images.<br />

1 To initiate image text entry, click on the blue bar under the image in the<br />

Imaging screen, Imaging Full screen or the Analysis screen.<br />

2 An Image Information screen will appear.<br />

3 The System Caption<br />

shows the image and<br />

data that currently<br />

resides on the image.<br />

4 The Preferred Caption<br />

allows for text entry to<br />

customize the image<br />

caption.<br />

5 The Additional<br />

Comments allows<br />

for text entry<br />

of comments, 1<br />

which appear<br />

only on the printed<br />

Colposcopy Report<br />

and Image Reports.<br />

6 Once the OK<br />

button is clicked,<br />

any user-specified<br />

caption would replace<br />

the text shown<br />

below the thumbnail<br />

image. Cancel will<br />

refuse any changes<br />

to the Image<br />

text box.<br />

2<br />

3<br />

4<br />

5<br />

6<br />

Imaging Screen<br />

Image Information Screen<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

4.7


Section IV<br />

IMAGING ■ Continued<br />

Imaging Screen continued<br />

Print Selected Image<br />

■<br />

■<br />

■<br />

Print Selected Image<br />

By Itself prints a large image (selected by the user in the Imaging screen)<br />

with lines for annotation.<br />

Print Selected Image With<br />

Blank Colposcopy Report prints<br />

a blank Colposcopy Report with<br />

the selected image. The report is<br />

then filled out by hand.<br />

You can create an Anonymous patient<br />

report by checking the box<br />

“Anonymous Patient”.<br />

Patient Reports<br />

Print Selected Image Screen<br />

We have provided two printed patient report examples for your review.<br />

The patient reports can be viewed in order on the following two pages.<br />

Report 1 ■ Blank Colposcopy Report<br />

This report with blank annotation lines can be printed by clicking the Print<br />

button and selected image “With Blank Colposcopy Report.” The last image<br />

selected on the Imaging screen will be the image contained in the report.<br />

Report 2 ■ Image-by-Itself Report<br />

This single, full screen image with annotation is printed by clicking on<br />

the print button first, then on the “By Itself” selection. Blank lines are<br />

printed automatically with the Image for physician annotation.<br />

End Visit Button Imaging Screen<br />

■<br />

■<br />

The End Visit button must be<br />

used to end the visit.<br />

A patient Auto Backup gauge will appear if the Auto Backup box<br />

is checked. This will back up the patient file to disk.<br />

4.8<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


IMAGING ■ Continued<br />

Section IV<br />

Imaging Screen continued<br />

Report #1<br />

Blank Colposcopy Report<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

4.9


Section IV<br />

IMAGING ■ Continued<br />

Imaging Screen continued<br />

Report #2<br />

Image-by-Itself Report<br />

4.10<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


IMAGING ■ Continued<br />

Section IV<br />

Imaging Screen continued<br />

Turning the System Off<br />

If you are finished with the Prism Digital Colposcopy System for<br />

the day, turn the system off.<br />

■<br />

■<br />

■<br />

■<br />

■<br />

Turn off the light source located on the lower Arm.<br />

Click on the Power Down button at the bottom of the Welcome screen.<br />

A dialog box will appear asking for confirmation that you wish to<br />

power down.<br />

Click on “OK” and wait for indication that it is now safe to turn off<br />

your computer.<br />

Turn the computer to off using the rocker switch on the computer.<br />

NOTE: The system can be left on all day without harm.<br />

Out of Paper<br />

■<br />

If you have a connection between the Prism Digital Colposcopy System<br />

and a printer, you will see the message below when the printer is out<br />

of paper.<br />

Load more paper into the printer and press the Resume button.<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

4.11


Section IV<br />

IMAGING ■ Continued<br />

Imaging Screen continued<br />

Ink Cartridges<br />

After changing ink cartridges you may need to align them.<br />

Follow the printer manufacturer’s instructions on the screen.<br />

NOTE: Thoroughly read the printer instruction manual before performing<br />

maintenance or operation of the device. Leisegang does not<br />

support peripheral devices such as printers.<br />

4.12<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


SYSTEM BACKUP & RESTORATION<br />

Section V<br />

System Backup & Restoration<br />

Using Automatic Backup<br />

Automatic Backup is used to create a copy of all patients files onto the<br />

Removable media Disk at the same time the data is copied to the hard drive.<br />

It is recommended that periodic storage to removable disks be done to<br />

safeguard files. This feature is not recommended with the Prism Digital<br />

Colposcopy System unless you have an external removable media drive<br />

attached to the Prism Digital Colposcopy System, or the optional Network<br />

Interface Software installed.<br />

Automatic Backup at Default setting is only accessible through the<br />

System Maintenance button on the Physician List screen.<br />

Click on the System Maintenance<br />

button on the Physician List screen.<br />

1 Click on the Disk Storage tab on the System Maintenance screen.<br />

2 Click on Automatic Backup when visit is closed.<br />

NOTE: A removable media drive must be present or an error message will appear.<br />

3 The gauge is a visual clue as to available memory in the Internal Disk<br />

or the Removable Disk.<br />

2<br />

1<br />

3<br />

NOTE: If a visit has been<br />

started, Automatic<br />

Backup may be<br />

accessed by clicking<br />

on Clinician Name<br />

button, followed by<br />

clicking on the<br />

System Maintenance<br />

button.<br />

System Maintenance Screen<br />

Disk Storage Tab<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

5.1


Section V<br />

SYSTEM BACKUP & RESTORATION ■ Continued<br />

System Backup continued<br />

When the box is checked,<br />

automatic backup occurs<br />

after clicking on the<br />

Yes or No button on the<br />

Print Today’s Colposcopy<br />

Report confirmation<br />

window. A Patient<br />

Automatic Backup window will appear.<br />

NOTE: A removable media drive must be present for this feature to work.<br />

Checking the Amount of Disk Space for<br />

Automatic Backup<br />

When a patient visit is started and the<br />

Automatic backup is enabled, a check is made<br />

that the removable disk is in the disk drive.<br />

If a removable disk is not in the drive, a<br />

window will appear indicating that the<br />

disk is not present.<br />

If a removable disk is in the disk drive, a<br />

second check is made to assure that<br />

adequate space for a patient visit is<br />

available on the disk. This check is<br />

an approximation.<br />

When the patient visit is closed, the actual<br />

amount of available space is determined.<br />

If the removable disk does not have<br />

sufficient space, a window will appear<br />

indicating that.<br />

Patient Auto Backup Window<br />

At this point, replace the removable<br />

disk with one that has available memory,<br />

and click on the Retry button. The same<br />

window will appear again if the new disk does not have sufficient space.<br />

If you don’t have a removable disk with a sufficient amount of space,<br />

click on the Cancel button.<br />

A window will appear indicating that you did not backup your patient visit.<br />

Click on the OK button.<br />

5.2<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


SYSTEM BACKUP & RESTORATION ■ Continued<br />

Section V<br />

System Backup continued<br />

Using Manual Backup for a Single Patient<br />

You can manually backup patient information from the Internal Disk<br />

to the Removable Disk only when a patient visit is closed.<br />

Go to the Patient screen and click<br />

on the Patient List button.<br />

1 Select a patient from the Patient List by clicking on her name, which will<br />

then be highlighted in yellow.<br />

2 With a patient’s name highlighted, click on the Backup button. The file<br />

has now been “backed-up.”<br />

1<br />

If there is not sufficient<br />

disk space available,<br />

a window will appear<br />

with a warning. If<br />

sufficient disk space is<br />

available, a Status bar<br />

will appear indicating<br />

that backup has<br />

begun, and will remain<br />

until completed.<br />

2<br />

Patient List Screen<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

5.3


Section V<br />

SYSTEM BACKUP & RESTORATION ■ Continued<br />

System Backup continued<br />

Using Manual Backup for a Group of Patients<br />

You can manually backup patient information from the Internal Disk<br />

to the Removable Disk only when a Patient Visit is closed.<br />

To select a group of patients for backup:<br />

Go to the Patient screen and click<br />

on the Patient List button.<br />

1 Select a group of patients from the patient list by clicking on a start point.<br />

With the mouse select button depressed, drag the Cursor over the group of<br />

patients until they are highlighted in yellow. The OK button will be deactivated.<br />

2 Click on the column headers to create a sort order you want the data to<br />

appear in. The sort order is revealed at the top right of the screen.<br />

3 Click on the Backup button to backup data to the Removable Disk<br />

or<br />

4 Click on Cancel to refuse the backup command.<br />

2<br />

1<br />

3<br />

4<br />

Patient List Screen<br />

5.4<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


SYSTEM BACKUP & RESTORATION ■ Continued<br />

Section V<br />

System Backup continued<br />

Using Manual Backup for All Patients<br />

You can manually backup patient information from the Internal Disk to the<br />

Removable Disk only when a Patient Visit is closed. To select all<br />

patients for backup:<br />

Go to the Patient screen and click<br />

on the Patient List button.<br />

1 Select all patients by clicking on the Select All button.<br />

2 Click on the column headers to create a sort order you want the data to<br />

appear in. The sort order is revealed at the top right of the screen.<br />

3 Click on the Backup button to backup data to the Removable Disk<br />

or<br />

4 Click on Cancel to refuse the backup command.<br />

1<br />

2<br />

3<br />

4<br />

Patient List Screen<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

5.5


Section V<br />

SYSTEM BACKUP & RESTORATION ■ Continued<br />

System Backup continued<br />

Using Manual Restore<br />

You can restore patient information from a Removable disk to the internal disk<br />

only when the patient visit is closed. Close the patient case by selecting the<br />

End Visit button in the Imaging screen.<br />

NOTE: A removable media disk must be present for this feature to work.<br />

Go to the Patient screen and click on<br />

the Patient List button.<br />

1 Select removable disk on the “Patient List From” combo box on the upper<br />

left corner of Patient List screen. You will now be able to see patient files<br />

stored on the removable disk.<br />

2 The Backup button will now say restore.<br />

3 Select a patient from the Patient List by clicking on her name, which<br />

will then be highlighted in yellow.<br />

■<br />

Click on the Restore button to copy a selected patient’s file back onto<br />

the Internal Disk.<br />

NOTE: To restore a group of patients or all patients to the Internal Drive, follow the same<br />

instruction for Patient backup listed on page 8.5.<br />

1<br />

3<br />

2<br />

Patient List Screen<br />

5.6<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


SYSTEM BACKUP & RESTORATION ■ Continued<br />

Section V<br />

System Backup continued<br />

A window will appear indicating the time it will take to copy the patient’s case<br />

to the Internal Disk. After the case has been restored, the patient’s name will<br />

appear in the Internal Disk Patient List.<br />

The Removable Disk Patient List does not update itself. You must reselect the<br />

Removable Disk from the top right combo box to remove the patient from the<br />

Removable Disk Patient List.<br />

If you attempt to restore a patient case<br />

from a disk that has been removed, or if<br />

you have a corrupt patient file, a window<br />

will appear explaining that the patient<br />

folder is missing.<br />

■<br />

■<br />

■<br />

To restore a patient file from a disk that has been removed, click on OK.<br />

Insert the Removable Disk into the disk drive.<br />

Select Removable Disk from the combo box at the top left of the screen<br />

to refresh the patient list, and retry.<br />

Printing from the Patient List<br />

■<br />

■<br />

■<br />

The list of patients on the removable disk can be viewed and printed<br />

at any time by first selecting removable disk from the combo box at<br />

the upper left of the Patient List screen.<br />

Select the Print button to initiate the<br />

print sequence.<br />

After selecting the Print button the<br />

following confirmation message appears.<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

5.7


Section V<br />

SYSTEM BACKUP & RESTORATION ■ Continued<br />

System Backup continued<br />

Deleting a Patient’s File from the Patient List<br />

A patient’s file (all visits) can be deleted from either the<br />

Internal or Removable Disk.<br />

NOTE: A removable media drive must be present to delete a file from a removable disk.<br />

1 Click on the patient’s name in the Patient List.<br />

2 Click on the Delete button at the bottom middle of the screen.<br />

(Remember, this deletes all visits for the patient.)<br />

■<br />

A Delete Patient Folder? window will appear.<br />

3 Click on the OK button if you want to delete the patient’s case, or click on<br />

the Cancel button if you do not.<br />

After the patient file is deleted, the Patient List screen will update itself.<br />

1<br />

2<br />

3<br />

Patient List Screen<br />

5.8<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


SYSTEM BACKUP & RESTORATION ■ Continued<br />

Section V<br />

System Backup continued<br />

Editing a Patient’s Name<br />

A patient’s name can be edited to reflect a name change.<br />

Click on the Edit Name button.<br />

1 The Edit Patient Name window will appear. Make any necessary edits to<br />

the patient name or ID.<br />

2 Click OK to set changes or Cancel to refuse changes and return to the<br />

Patient List screen.<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

5.9


Section V<br />

Notes<br />

5.10<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


SYSTEM MAINTENANCE ■ Continued<br />

Section VI<br />

System Maintenance<br />

Accessing System Maintenance<br />

The System Maintenance screen allows you to add your facility name,<br />

enable or disable security passwords, access details to the System, select<br />

a printer, evaluate disk space, self check the camera, and access the<br />

Windows 95 operating system.<br />

You can access System Maintenance through the Attending Clinician List<br />

screen during a visit by clicking on the Attending Clinician Name button seen<br />

in the Colposcopy Report or the Imaging screen, followed by the System<br />

Maintenance button.<br />

Click on the System Maintenance button<br />

on the Attending Clinician List screen.<br />

Attending Clinician List Screen<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

6.1


Section VI<br />

SYSTEM MAINTENANCE ■ Continued<br />

System Maintenance continued<br />

Setting up User Default Settings<br />

User defaults allow you to customize some of the data fields as they<br />

appear when you initially access them.<br />

■<br />

■<br />

■<br />

■<br />

■<br />

■<br />

Click on the User Default tab.<br />

Select the default setting you want to appear in the Colposcopy Report<br />

sections prior to entering data.<br />

The default options for the Colposcopy Report are specific to the<br />

History and Risks screens.<br />

The Contrast button from the Imaging section can also be user<br />

defaulted to Full Contrast or No Contrast.<br />

The Brightness mode is used for systems where the shutters<br />

need to be disabled.<br />

The Brightness/Contrast User Default are specific to the Focus Mode<br />

feature. Adjusting the settings to Contrast and Brightness will create<br />

a User Default setting. Clicking on the Factory Default will revert the<br />

User setting to the Factory Default standard.<br />

System Maintenance Screen<br />

User Default Tab<br />

6.2<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


SYSTEM MAINTENANCE ■ Continued<br />

Section VI<br />

System Maintenance continued<br />

Entering Facility Information<br />

When you print the Image Reports, the facility information appears<br />

as the header.<br />

■<br />

■<br />

Click on the Facility tab.<br />

Type your facility contact information in the blank spaces.<br />

To get to the next data field, use the Tab button on your keyboard.<br />

System Maintenance Screen<br />

Facility Tab<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

6.3


Section VI<br />

SYSTEM MAINTENANCE ■ Continued<br />

System Maintenance continued<br />

Accessing Various Aspects of the System<br />

The Computer System screen gives you access and selective control over<br />

various functions of the Windows 95 platform, which controls the peripheral<br />

devices and hard drive management function. You may control the<br />

following functions from the Computer System screen:<br />

System Maintenance Screen<br />

Computer System Tab<br />

■<br />

■<br />

■<br />

■<br />

■<br />

■<br />

Changing the<br />

date and time.<br />

Changing mouse<br />

properties.<br />

Scanning a disk<br />

to check for errors<br />

or viruses.<br />

Defragmenting a<br />

disk drive.<br />

Setting up the printer.<br />

Accessing the print<br />

queue.<br />

Changing the Date and Time<br />

■<br />

■<br />

■<br />

■<br />

■<br />

■<br />

Click on the Computer System tab.<br />

Click on the Date/Time button in<br />

the System section.<br />

Click on the appropriate date in<br />

the calendar to change the day.<br />

Click on the arrow to the right of<br />

the month to change the month.<br />

Click on the arrows to the right of<br />

the year to change the year.<br />

Click on the arrows to the right<br />

of the time to set the time.<br />

System Maintenance Screen<br />

Computer System Tab<br />

Date/Time Properties<br />

NOTE: Further access to the Windows 95 environment is not recommended. Inadvertent tampering<br />

with Windows 95 can change the performance of the Prism software.<br />

6.4<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


SYSTEM MAINTENANCE ■ Continued<br />

Section VI<br />

System Maintenance continued<br />

Changing the Time Zone<br />

■<br />

■<br />

■<br />

Click on the arrow to the right of the Time Zone to change<br />

the time zone.<br />

Click on the Apply button to apply all of your changes to the system.<br />

Click on the OK button to return to the Computer System screen.<br />

System Maintenance Screen<br />

Computer System Tab<br />

Time Zone Adjustment<br />

Changing Mouse Properties<br />

■ Click on the Button section to change clicking speed and operations.<br />

■ Click on the Pointer section to change cursors.<br />

■ Click on the Motion button to change the cursor movement.<br />

Mouse Button Screen<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

6.5


Section VI<br />

SYSTEM MAINTENANCE ■ Continued<br />

System Maintenance continued<br />

Scanning a Disk<br />

to Check for Errors or Viruses<br />

■<br />

Click on Scan Disk to check the<br />

hard drive or Removable drive.<br />

Scan Disk<br />

Defragmenting<br />

a Disk Drive<br />

■<br />

Click on Defragment<br />

Disk to defragment<br />

the hard drive or<br />

Removable drive.<br />

System Maintenance Screen ■ Computer Tab<br />

Printer Set-up<br />

Screens<br />

■<br />

For detailed<br />

information on your<br />

printer and how to<br />

operate it, refer to<br />

your printer’s<br />

Operating Manual.<br />

Printer Set-Up Screen<br />

6.6<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System


SYSTEM MAINTENANCE ■ Continued<br />

Section VI<br />

System Maintenance continued<br />

NOTE: The screens below may show slightly different data depending on the printer<br />

model you are using.<br />

Printer Maintenance Screens<br />

Checking the Amount of Space on a Disk<br />

The Disk Storage tab shows the current disk utilization of the internal<br />

and removable disks in numerical and graphical form. This information<br />

is updated continuously.<br />

■<br />

Click on the Disk Storage tab.<br />

The top gauge indicates<br />

how much internal disk<br />

space is used and how<br />

much is available. The<br />

bottom gauge indicates<br />

how much removable<br />

disk space is used and<br />

how much is available.<br />

A removable media<br />

drive must be present<br />

for this feature to work.<br />

System Maintenance Screen<br />

Disk Storage Tab<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System<br />

6.7


Section VI<br />

Notes<br />

6.8<br />

User’s Guide<br />

Leisegang Prism Digital Colposcopy System