800 rpm max. Step 3 – Mark <strong>the</strong> implant axis<br />

With <strong>the</strong> Ø 2.2 mm pilot drill, mark <strong>the</strong> implant axis by<br />

drilling to a depth of about 6 mm.<br />

Insert <strong>the</strong> short side of <strong>the</strong> depth gauge with <strong>the</strong> distance<br />

indicator to check for correct implant axis orientati<strong>on</strong>.<br />

If necessary, correct unsatisfactory implant axis orientati<strong>on</strong><br />

in <strong>the</strong> following step.<br />

Note<br />

The distance indicator visualizes <strong>the</strong> shoulder diameter of<br />

4.8 mm (RN) and enables checking of <strong>the</strong> probable positi<strong>on</strong><br />

of <strong>the</strong> implant shoulder.<br />

4<br />

800 rpm max. Step 4 – Prepare <strong>the</strong> implant bed to Ø 2.2 mm<br />

Pre-drill <strong>the</strong> implant bed to <strong>the</strong> final preparati<strong>on</strong> depth<br />

with <strong>the</strong> Ø 2.2 mm pilot drill.<br />

Use <strong>the</strong> Ø 2.2 mm alignment pin to check <strong>the</strong> implant<br />

axis and preparati<strong>on</strong> depth.<br />

Cauti<strong>on</strong><br />

At this point take an X-ray, particularly with vertically reduced<br />

b<strong>on</strong>e availability. The alignment pin is inserted into<br />

<strong>the</strong> drilled area, which allows a comparative visualizati<strong>on</strong><br />

of <strong>the</strong> drill hole in relati<strong>on</strong> to <strong>the</strong> anatomical structures.<br />

5<br />

600 rpm max. Step 5 – Widen <strong>the</strong> implant bed to Ø 2.8 mm<br />

C<strong>on</strong>tinue with <strong>the</strong> implant bed preparati<strong>on</strong>.<br />

If necessary, correct <strong>the</strong> implant positi<strong>on</strong> with <strong>the</strong><br />

Ø 2.8 mm pilot drill. Use <strong>the</strong> Ø 2.8 mm depth gauge<br />

to check <strong>the</strong> preparati<strong>on</strong> depth.<br />

For an implant with an endosteal diameter of 3.3 mm,<br />

basic preparati<strong>on</strong> ends here. C<strong>on</strong>tinue with <strong>the</strong> fine implant<br />

bed preparati<strong>on</strong> <strong>on</strong> page 37.<br />

34<br />

4. Surgical <strong>procedures</strong> 4.1 Implant bed preparati<strong>on</strong>